# Accurate and Sensitive UHPLC–Tandem Mass Spectrometry Sequential Methods for Therapeutic Drug Monitoring of Aztreonam/Avibactam in Human Plasma

**Authors:** Ilaria Trozzi, Beatrice Giorgi, Riccardo De Paola, Milo Gatti, Federico Pea

PMC · DOI: 10.3390/pharmaceutics18030377 · Pharmaceutics · 2026-03-19

## TL;DR

This paper presents a new method to accurately measure drug levels in patients using advanced mass spectrometry, helping doctors adjust doses for better treatment outcomes.

## Contribution

The study introduces validated UHPLC–qTOF MS/MS methods for precise TDM of aztreonam and avibactam in human plasma.

## Key findings

- The methods showed excellent linearity (R² ≥ 0.999) for aztreonam and avibactam in plasma.
- Lower limits of quantification were 0.2 µg/mL for aztreonam and 0.1 µg/mL for avibactam.
- Clinical application revealed significant inter-individual variability in drug concentrations.

## Abstract

Background/Objectives: The aztreonam/avibactam combination represents a promising therapeutic option for severe infections caused by multidrug-resistant Gram-negative pathogens, particularly in critically ill patients. Due to marked pharmacokinetic variability and the need to achieve joint pharmacokinetic/pharmacodynamic (PK/PD) targets of both agents, therapeutic drug monitoring (TDM) may play a pivotal role in optimizing treatment. This study aimed to develop and validate two rapid, accurate, and sensitive UHPLC–qTOF MS/MS sequential methods for quantifying aztreonam and avibactam in human plasma, suitable for routine clinical TDM. Methods: Plasma concentrations were determined by means of ultra-high-performance liquid chromatography coupled with quadrupole time-of-flight tandem mass spectrometry (UHPLC–qTOF MS/MS), operating in positive and negative electrospray ionization modes for aztreonam and avibactam, respectively. Sample preparation consisted of protein precipitation with isotopically labeled internal standards. The method’s validation was performed according to the European Medicines Agency guidelines, by assessing selectivity, linearity, precision, accuracy, recovery, matrix effects, carry-over, and stability. Clinical applicability was evaluated by reprocessing plasma samples, which were already previously collected for routine clinical practice from 20 hospitalized patients undergoing treatment with ceftazidime-avibactam plus aztreonam. Results: The methods showed excellent linearity (R2 ≥ 0.999) over ranges of 0.2–100 µg/mL for aztreonam and 0.1–50 µg/mL for avibactam. Lower limits of quantification were 0.2 µg/mL and 0.1 µg/mL, respectively. Intra- and inter-day precision and accuracy met the EMA criteria at all of the quality control levels. Extraction recovery exceeded 90% for both analytes, and matrix effects were effectively compensated by internal standards. Stability testing highlighted the need for careful sample handling, particularly for aztreonam under repeated freeze–thaw conditions. Clinical application revealed substantial inter-individual variability in steady-state concentrations. Conclusions: The validated UHPLC–qTOF MS/MS assays provide robust and sensitive sequential quantification of aztreonam and avibactam in human plasma, supporting TDM-guided dose optimization in clinical practice.

## Linked entities

- **Chemicals:** aztreonam (PubChem CID 5742832), avibactam (PubChem CID 9835049)

## Full-text entities

- **Diseases:** infections (MESH:D007239)
- **Chemicals:** ceftazidime-avibactam (MESH:C000595613), aztreonam (MESH:D001398), Aztreonam/Avibactam (-), avibactam (MESH:C543519)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC13029189/full.md

## Figures

6 figures with captions in the complete paper: https://tomesphere.com/paper/PMC13029189/full.md

## References

33 references — full list in the complete paper: https://tomesphere.com/paper/PMC13029189/full.md

---