Mechanical performance of narrow dental implants with varying lengths and angled abutments
Darshana Raut, Gaurav Beohar, Prince Kumar, Mayur Khairnar, Babita Pawar, Nilesh Fatale

TL;DR
This study compares the mechanical strength of narrow dental implants with different lengths and angled abutments under static and fatigue loads.
Contribution
The study reveals that longer implants have higher static strength while shorter ones perform better under fatigue.
Findings
Group 1 (3.5 mm x 10 mm, 25° abutment) showed 418 N static and 182 N fatigue resistance.
Group 2 (3.5 mm x 25 mm, 45° abutment) showed 454 N static and 160 N fatigue resistance.
Shorter implants performed better under fatigue conditions compared to longer ones.
Abstract
Mechanical performance of narrow-diameter dental implants with varying lengths and abutment angulations remains a critical factor in long-term clinical success. Therefore, it is of interest to evaluate Quickdent implant systems under static and fatigue loading according to ISO 14801:2016E. Group 1 (3.5 mm x 10 mm, 25° abutment) and Group 2 (3.5 mm x 25 mm, 45° abutment) were tested using fifteen assemblies per group. Group 1 showed 418 N static and 182 N fatigue resistance, while Group 2 showed 454 N static and 160 N fatigue resistance with distinct failure modes. Thus, we show that longer implants offer higher static strength, whereas shorter ones perform better under fatigue conditions.
Genes, proteins, chemicals, diseases, species, mutations and cell lines named across the full text — each resolved to its canonical identifier and authoritative record.
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Taxonomy
TopicsDental Implant Techniques and Outcomes · Dental materials and restorations · Scientific and Engineering Research Topics