Diagnostic Performance of Leukocyte Abnormality Detection in a Large Cohort of Healthy Blood Donors Using Sysmex XN Series Analyzers Integrated with Peripheral Blood Morphology and Flow Cytometry
Francesca Romano, Valentina Becherucci, Sara Ciullini Mannurita, Edda Russo, Alessandra Mongia, Anna Maria Grazia Gelli, Alessandra Fanelli, Francesca Brugnolo

TL;DR
This study evaluates how well the Sysmex XN series detects leukocyte abnormalities in healthy blood donors, comparing automated analysis with manual and flow cytometry methods.
Contribution
The study introduces a comprehensive evaluation of the diagnostic performance of the Sysmex XN series in detecting leukocyte abnormalities in a large cohort of healthy donors.
Findings
WPC analysis provided higher specificity than WDF alone in detecting leukocyte abnormalities.
Approximately 50% of flagged cases showed reactive morphological features confirmed by flow cytometry.
Transient immune activation, possibly due to vaccination, influenced leukocyte flagging.
Abstract
Background: The Sysmex XN series (XN-1000 and XN-9100, Sysmex Corporation, Kobe, Japan) represents a latest-generation automated hematology platform integrating fluorescence-based technologies and multi-channel analysis (WDF and WPC) to improve leukocyte characterization. This study aimed to evaluate the performance of the Sysmex XN series in detecting leukocyte abnormalities flagged during routine complete blood count analysis in a large cohort of healthy donors, using morphological assessment and flow cytometry as confirmatory methods. Methods: Approximately 8000 healthy blood donors from the AOU Meyer Transfusion Centre were evaluated between 2021 and 2024. All samples underwent CBC analysis using the XN-1000 and XN-9100 analyzers with the WDF channel. Samples showing WBC-related flags were subjected to reflex testing with the WPC channel, followed by digital blood smear review using…
Genes, proteins, chemicals, diseases, species, mutations and cell lines named across the full text — each resolved to its canonical identifier and authoritative record.
Click any figure to enlarge with its caption.
Figure 1
Figure 2
Figure 3
Figure 4Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Taxonomy
TopicsDigital Imaging for Blood Diseases · Biosensors and Analytical Detection · Clinical Laboratory Practices and Quality Control