# Clinical Evaluation of a Novel Point‐of‐Care Hematology Analyzer for Complete Blood Count With Differential

**Authors:** Ryan C. Shean, Sterling T. Bennett

PMC · DOI: 10.1111/ijlh.70032 · International Journal of Laboratory Hematology · 2025-12-08

## TL;DR

A new point-of-care blood analyzer was tested in various clinical settings and showed mostly good agreement with lab results, though some differences were found.

## Contribution

The study evaluates a novel point-of-care hematology analyzer's performance in real-world clinical settings.

## Key findings

- HemoScreen showed strong correlations for HGB, RBC, PLT, WBC, and neutrophil count with central lab results.
- HCT and MCV values were significantly lower on HemoScreen compared to the central lab system.
- Neutrophil percentage was higher, while monocyte counts were significantly lower on HemoScreen.

## Abstract

Point‐of‐care testing (POCT) for hematology enables clinicians to access actionable results during patient visits, supporting timely decisions such as transfusions or chemotherapy adjustments. However, POCT analyzers must demonstrate clinical performance comparable to centralized laboratory systems. We conducted a proof‐of‐concept evaluation of the HemoScreen (PixCell Medical, Israel) hematology POCT analyzer across diverse clinical settings in an integrated healthcare system.

EDTA whole blood specimens from inpatient wards, emergency departments, outpatient clinics, and infusion centers underwent routine CBC testing on a Sysmex XN analyzer (Sysmex Corporation, Japan) at the central laboratory. Residual, de‐identified samples were tested within 10 min on a HemoScreen analyzer. All parameters and instrument flags were recorded.

From 9/15/2020 to 11/20/2020, 199 samples were analyzed, including 49 from the emergency department and 50 each from inpatient, outpatient, and infusion center settings. HGB, RBC, PLT, WBC, and absolute neutrophil count showed strong linear correlations (R
2 = 0.91–0.98) and no significant differences between instruments (p > 0.29). HCT and MCV values were significantly lower on HemoScreen (bias −5% and −3.6%; p = 0.010 and p < 0.001, respectively), with modestly reduced correlations (R
2 = 0.80–0.89). Neutrophil percentage was higher on HemoScreen (bias +9.0%, p < 0.001), whereas monocyte counts were significantly lower (bias −50%, p < 0.001).

HemoScreen demonstrated generally acceptable agreement with the Sysmex XN for most parameters. However, systematic differences in HCT, MCV, neutrophil, and monocyte results warrant further investigation to assess clinical impact before broader implementation.

## Full-text entities

- **Chemicals:** EDTA (MESH:D004492)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12956494/full.md

## Figures

2 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12956494/full.md

## References

26 references — full list in the complete paper: https://tomesphere.com/paper/PMC12956494/full.md

---
Source: https://tomesphere.com/paper/PMC12956494