# Knowledge based automated planning for non‐coplanar VMAT stereotactic radiosurgery in ocular malignancies

**Authors:** Chase Cochran, Shane McCarthy, William St. Clair, Damodar Pokhrel

PMC · DOI: 10.1002/acm2.70530 · Journal of Applied Clinical Medical Physics · 2026-03-03

## TL;DR

This study shows that automated planning using RapidPlan with HyperArc improves efficiency and quality in radiosurgery for eye cancers.

## Contribution

A novel HARP model combining RapidPlan and HyperArc is introduced for efficient and consistent ocular radiosurgery planning.

## Key findings

- Model-based plans improved OAR sparing with lower doses to the optic nerve and lens.
- Planning time was reduced from 120 minutes to 15 minutes using the HARP model.
- QA and MC second-check results improved with automated planning.

## Abstract

Stereotactic radiosurgery/radiotherapy (SRS/SRT) has emerged as a less invasive alternative to enucleation in the management of ocular malignancies. SRS/SRT planning is time‐consuming, complex, and the plan quality depends on the experience of the treatment planner. We demonstrate the use of RapidPlan modeling in conjunction with HyperArc for the treatment of ocular diseases to improve planning efficiency, consistency, and quality.

A HyperArc‐based RapidPlan (HARP) model was iteratively trained on 80 patient datasets that simulated ocular malignancies. Twenty additional patient datasets were reserved for model testing and comparison with manual plans. Each testing dataset was both manually planned and replanned with the final RapidPlan model. Target volumes were defined in the HyperArc module as the GTV and PTV. The PTV was generated by a 2 mm static expansion of the GTV. 25 Gy in 1 fraction was prescribed, and all plans were normalized such that PTVD95% = 25 Gy. Treatment planning was done using Eclipse V16 with the Acuros XB dose engine on a TrueBeam LINAC with Millennium 120 MLCs (5 mm width). All plans underwent EPID‐based patient‐specific QA and an independent Monte Carlo (MC) second‐check.

Model‐based plans demonstrated similar degrees of conformality, gradient, and target coverage compared to manual planning. OAR sparing showed statistically significant improvements with model‐based plans. Optic nerve Dmax (D0.03 cc) decreased to 4.56 Gy (manual: 7.18 Gy), and lens Dmax decreased to 7.64 Gy (manual: 9.36 Gy). Lacrimal gland mean dose decreased from 11.61 to 6.47 Gy. Modulation factor, monitor units, and beam‐on times also all decreased. Patient‐specific QA results showed improvements compared to manually generated plans. MC second check results also improved, increasing from an average of 97.81%–98.74%. Plan optimization times decreased significantly from approximately 120 min for manual planning to 15 min on average for model‐based plans.

RapidPlan‐generated ocular SRS plans showed either comparable performance or improvements in all plan metrics measured, compared to manual planning. Planning times were significantly reduced while maintaining or improving plan quality. Clinical implementation of this HARP model is ongoing at our clinic.

## Full-text entities

- **Diseases:** ocular disease (MESH:D005128), brain malignancies (MESH:D001932), HARP (MESH:D019292), RP (MESH:D012174), ocular/intraocular lesions (MESH:D064090), Tumor (MESH:D009369), SRS (MESH:C536678)
- **Chemicals:** gadolinium (MESH:D005682), Iodine (MESH:D007455), OAR (-), Ruthenium (MESH:D012428)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12956477/full.md

## Figures

3 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12956477/full.md

## References

44 references — full list in the complete paper: https://tomesphere.com/paper/PMC12956477/full.md

---
Source: https://tomesphere.com/paper/PMC12956477