# Evaluation of the Dose Distribution Robustness for Lung Tumor Tracking in Robotic Radiotherapy Using Four-Dimensional Computed Tomography

**Authors:** Yasuhide Miyabe, Hiromu Yamanaka, Hiroto Seki, Hiromitsu Iwata, Jyunetsu Mizoe

PMC · DOI: 10.7759/cureus.102341 · Cureus · 2026-01-26

## TL;DR

This study evaluates how respiratory motion affects dose distribution in lung tumor radiotherapy using 4DCT simulations to improve treatment planning accuracy.

## Contribution

The study introduces a feasibility method using 4DCT to assess dose variations caused by respiratory phase changes in CyberKnife treatments.

## Key findings

- Patient 1 experienced CTV dose reductions below prescribed levels at an inspiratory phase.
- Patient 2 showed minor dose reductions but maintained prescribed dose coverage.
- No clinically relevant dose deterioration was observed in OARs for either patient.

## Abstract

Respiratory motion of tumors and surrounding organs at risk (OARs) is a critical factor affecting dose distribution and treatment accuracy in radiotherapy. The CyberKnife Synchrony respiratory tracking system (Accuray Inc., Sunnyvale, CA, USA) improves targeting accuracy by estimating tumor position based on the correlation between implanted fiducial markers and external respiratory motion. However, in clinical practice, dose evaluation is commonly performed using computed tomography (CT) images acquired at a single respiratory phase, and the impact of respiratory phase variations on dose distribution has not been fully investigated.

This study is a feasibility analysis aimed at demonstrating the potential utility of four-dimensional CT (4DCT)-based simulations for evaluating dose variations associated with respiratory phase differences during Synchrony-based treatments. Two patients with lung tumors treated using CyberKnife stereotactic radiotherapy (SRT) were included in the study. Treatment plans (reference plans) were created using breath-hold CT images acquired at the end-expiratory phase. Using the same beam parameters as the reference plans, dose calculations were performed on multiple respiratory-phase images derived from free-breathing 4DCT acquired during treatment planning.

Dose-volume histogram (DVH) metrics, including clinical target volume (CTV) D99 and planning target volume (PTV) D95, as well as dose parameters for OARs, were compared with the clinical goals defined by the radiation oncologist. In Patient 1, dose reductions were observed in both the PTV and CTV at a specific inspiratory phase, resulting in the CTV dose falling below the prescribed level. In contrast, Patient 2 showed only minor dose reductions in the PTV and CTV, while prescribed dose coverage was maintained. No clinically relevant dose deterioration was observed for the OARs in either patient.

The proposed evaluation method using 4DCT demonstrated the potential to identify the influence of respiratory phase variations on dose distribution in Synchrony-based respiratory tracking radiotherapy at the treatment planning stage. This approach may serve as a supplementary tool to conventional single-phase CT-based dose evaluation for assessing plan robustness against respiratory motion.

## Full-text entities

- **Diseases:** Lung Tumor (MESH:D008175), tumor (MESH:D009369), adenoid cystic carcinoma (MESH:D003528), lung metastases (MESH:D009362)
- **Chemicals:** gold (MESH:D006046), water (MESH:D014867), isodose (-)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12935493/full.md

## Figures

2 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12935493/full.md

## References

19 references — full list in the complete paper: https://tomesphere.com/paper/PMC12935493/full.md

---
Source: https://tomesphere.com/paper/PMC12935493