# Improving magnetic field homogeneity in prostate MR imaging and spectroscopy using an add-on local external shim coil array

**Authors:** Angeliki Stamatelatou, Carlijn Tenbergen, Sahar Nassirpour, Paul Chang, Jack van Asten, Arend Heerschap, Tom Scheenen

PMC · DOI: 10.1007/s10334-025-01290-y · Magma (New York, N.y.) · 2025-09-11

## TL;DR

A new external coil array improves magnetic field homogeneity in prostate MRI, leading to better image quality and spectroscopy precision.

## Contribution

A custom 16-channel external shim coil array is introduced to enhance B0 homogeneity in prostate MRI.

## Key findings

- The Dice similarity coefficient improved from 0.90 to 0.92 with the new coil array.
- The standard deviation of ΔB0 decreased from 18.5 Hz to 16.5 Hz.
- Citrate linewidths in MRSI were reduced from 8.3 Hz to 7.9 Hz.

## Abstract

To improve B0 field homogeneity in prostate MR imaging and spectroscopy using a custom-designed 16-channel external local shim coil array.

In vivo prostate imaging was performed in seven healthy volunteers (mean age: 40.7 years) without bowel preparation. A simulation study determined the most robust configuration of the add-on shim coil together with shim coils from the MR system. The combination of the add-on shim coil array with 1st-order shims of the MR scanner was used and compared to the MR system’s reference standard of conventional 1st and 2nd-order shimming. Performance was evaluated using the standard deviation (SD) of off-resonance frequencies (ΔB0) from B0-maps, the distortion of diffusion-weighted images (DWI) by the Dice similarity coefficient (DSC) with reference to T2W images, and the citrate linewidths of 1H-MR spectroscopic imaging (1H-MRSI) of the prostate.

Overall, improvements were made with the combination of coils, with DSC rising from 0.90 ± 0.02 to 0.92 ± 0.02 (p = 0.01), a reduction in the SD of ΔB0 from 18.5 ± 4.7 Hz to 16.5 ± 3.6 Hz, and a decrease in citrate linewidths from 8.3 ± 5.7 Hz to 7.9 ± 3.6 Hz (p = 0.018), all compared to the reference standard of conventional 1st and 2nd-order shimming.

A dedicated external shim coil array can improve B0 homogeneity in the prostate, enhancing image quality and precision in multi-parametric MRI.

The online version contains supplementary material available at 10.1007/s10334-025-01290-y.