# Understanding the mismatch between in-vivo and in-silico rhinomanometry

**Authors:** Marco Atzori, Gabriele Dini Ciacci, Maurizio Quadrio

PMC · DOI: 10.1007/s11517-025-03450-7 · Medical & Biological Engineering & Computing · 2025-09-25

## TL;DR

This study investigates why simulated and real-world nasal resistance measurements often disagree, finding that device design, specifically probe placement, may be a key cause.

## Contribution

The paper identifies a previously overlooked design bias in rhinomanometers due to probe placement affecting external pressure measurements.

## Key findings

- The position of the pressure tap in clinical devices significantly impacts measurement accuracy.
- Device design assumptions about flow development may introduce systematic bias in nasal resistance measurements.
- Numerical simulation uncertainties are minor compared to device-related discrepancies.

## Abstract

Numerical simulations and clinical measurements of nasal resistance are in quantitative disagreement. The order of magnitude of this mismatch, that sometimes exceeds 100%, is such that known sources of uncertainty cannot explain it. The goal of the present work is to examine a source of bias introduced by the design of medical devices, which has not been considered until now as a possible explanation. We study the effect of the location of the probe on the rhinomanometer that is meant to measure the ambient pressure. Rhinomanometry is carried out on a 3D silicone model of a patient-specific anatomy; a clinical device and dedicated sensors are employed side-by-side for mutual validation. The same anatomy is also employed for numerical simulations, with approaches spanning a wide range of fidelity levels. We find that the intrinsic uncertainty of the numerical simulations is of minor importance. To the contrary, the position of the pressure tap intended to acquire the external pressure in the clinical device is crucial, and can cause a mismatch comparable to that generally observed between in-silico and in-vivo rhinomanometry data. A source of systematic bias may therefore exist in rhinomanometers, designed under the assumption that measurements of the nasal resistance are unaffected by the flow development within the instruments.

## Full-text entities

- **Chemicals:** silicone (MESH:D012828)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12868048/full.md

## Figures

10 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12868048/full.md

## References

3 references — full list in the complete paper: https://tomesphere.com/paper/PMC12868048/full.md

---
Source: https://tomesphere.com/paper/PMC12868048