# P-2007. Accuracy and Precision of Self-Administered Audiograms Using a Smartphone Application

**Authors:** Nabin K Shrestha, Tricia Bravo, Omar Mehkri, Carmen Jamis

PMC · DOI: 10.1093/ofid/ofaf695.2171 · 2026-01-11

## TL;DR

This study evaluates a smartphone app for self-administered hearing tests, finding it accurate and precise compared to formal audiograms.

## Contribution

The study demonstrates that a smartphone app can provide precise and accurate hearing measurements comparable to formal audiograms.

## Key findings

- Self-administered audiograms using the Mimi Hearing App had a root mean square error of less than 5 dB at all tested frequencies.
- The mean error compared to formal audiograms ranged from -7.07 to 4.73 dB across tested frequencies.

## Abstract

Risk of hearing loss is a major concern for infectious disease physicians treating patients with prolonged courses of potentially ototoxic antimicrobials. Hearing loss can be detected by audiologic testing, but the logistics of obtaining formal audiograms are often challenging and obtaining serial evaluations for patients on prolonged treatments even more so. The Mimi Hearing App is a smartphone application that allows one to obtain self-administered audiograms. The purpose of this study was to evaluate the accuracy and precision of self-administered audiograms using this smartphone application.Table 1.Precision of measurement of hearing loss by self-administered audiograms using the Mimi Hearing App on repeat testing

Precision of measurement of hearing loss by self-administered audiograms using the Mimi Hearing App on repeat testing

Personnel working in the department of infectious diseases at Cleveland Clinic who had iPhones, and wired or wireless headphone devices compatible with iPhones, were invited to participate in the study. Each study subject was asked to record 10 self-administered audiograms on their iPhone, on 10 separate days, using the Mimi Hearing Application, and on completion of these, have a formal audiogram done by a Cleveland Clinic audiologist. Accuracy and precision of the self-administered audiograms at 250, 500, 1000, 2000, 4000, and 8000 Hz, were evaluated using the formal audiogram result as the gold standard.Table 2.Accuracy of measurement of hearing loss by self-administered audiograms using the Mimi Hearing App compared to formal audiologic assessments

Accuracy of measurement of hearing loss by self-administered audiograms using the Mimi Hearing App compared to formal audiologic assessments

Sixty self-administered audiograms were available from 6 subjects. One subject was in the age range 30-39, two in the range 40-49, two in the range 50-59, and one in the range 60-69. Three were male and three were female. Summarized audiograms from the 10 readings for each subject, depicting mean readings with standard error bars at each tested frequency, are shown in Figure 1. The measurements from the self-administered audiograms were very precise, with a root mean square error of less than 5 dB at each tested frequency (Table 1). When compared to measurements on formal audiologic testing, the mean error ranged from -7.07 to 4.73 dB for the various tested frequencies (Table 2).

Self-administered audiograms using the Mimi Hearing Application on a iPhone provide accurate measurements of hearing acuity with a high degree of precision between measurements.

All Authors: No reported disclosures