# Safety and Efficacy of the LVIS EVO Device for Stent-Assisted Coiling of Intracranial Aneurysms: A Systematic Review and Meta-Analysis

**Authors:** Abdulrahim Saleh Alrasheed, Amna Mutasim Elazrag, Rola Hamad Alseghair, Majd Nouh Alasmari, Mohammad Salem Alqahtani, Hosam Al-Jehani

PMC · DOI: 10.3390/jcm15010260 · Journal of Clinical Medicine · 2025-12-29

## TL;DR

This study reviews the safety and effectiveness of the LVIS EVO stent for treating brain aneurysms using coiling, finding it to be highly successful with low complication rates.

## Contribution

The study provides a systematic review and meta-analysis of the LVIS EVO stent's performance in stent-assisted coiling for intracranial aneurysms.

## Key findings

- Immediate angiographic results showed 78% complete aneurysm occlusion.
- Procedural and clinical complication rates were 5% and 6%, respectively.
- Follow-up showed 73% complete aneurysm obliteration.

## Abstract

Background: The low-profile visualized intraluminal support EVO (LVIS EVO) stent is a novel device for stent-assisted coiling (SAC) that offers improved visibility, radial force, and navigability in tortuous cerebral vessels. This systematic review and meta-analysis evaluated the safety and efficacy of the LVIS EVO stent in treating intracranial aneurysms (IAs). Methods: A systematic search of PubMed, Web of Science, Cochrane Library, and Scopus databases to identify studies reporting clinical outcomes of LVIS EVO in SAC of IAs was performed. Primary outcomes included aneurysm occlusion based on the Raymond–Roy Occlusion Classification (RROC), procedural, and clinical complication rates. Pooled estimates with 95% confidence intervals (CIs) were calculated using either a fixed- or random-effects model. Heterogeneity was assessed using the I2 statistic. Sensitivity analyses were conducted using the leave-one-out method. Results: Twelve studies involving 567 patients and 586 aneurysms were included. Immediate angiographic results showed complete occlusion (RROC I) in 78% (95% CI: 0.53–0.92), residual neck (RROC II) in 13% (95% CI: 0.08–0.21), and residual aneurysm (RROC III) in 9% (95% CI: 0.02–0.34). At follow-up, complete obliteration was observed in 73% (95% CI: 0.60–0.83). Procedural complications occurred in 5% (95% CI: 0.03–0.09), and clinical complications in 6% (95% CI: 0.04–0.10). Conclusions: The LVIS EVO stent demonstrated high technical success, effective aneurysm occlusion, and low complication rates, supporting its safety and efficacy. Prospective studies are warranted to confirm long-term outcomes.

## Full-text entities

- **Diseases:** IAs (MESH:D002532), aneurysm (MESH:D000783)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12786705/full.md

## Figures

8 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12786705/full.md

## References

37 references — full list in the complete paper: https://tomesphere.com/paper/PMC12786705/full.md

---
Source: https://tomesphere.com/paper/PMC12786705