# Performance evaluation of the nanoScan® P123S total-body PET

**Authors:** Dániel Réti, Carlos-Alcaide Corral, Islay Cranston, Victoria J. M. Reid, Kerry M. O’Rourke, Timaeus E. F. Morgan, Axel Montagne, Maurits A. Jansen, Valeria K. Burianova, Andrew Sutherland, Péter Major, Kálmán Nagy, Gergő Bagaméry, Adriana A. S. Tavares

PMC · DOI: 10.1186/s40658-025-00817-5 · 2025-12-08

## TL;DR

This paper evaluates a new PET scanner for total-body imaging in rodents, showing improved performance in sensitivity and resolution.

## Contribution

The study provides a detailed performance evaluation of the nanoScan® P123S PET system for total-body preclinical imaging.

## Key findings

- The scanner achieved a spatial resolution under 0.7 mm with iterative reconstruction.
- The NECR peak was 1805.0 kcps at 93.7 MBq for mouse-sized phantoms.
- Total-body rat images were successfully acquired using the new system.

## Abstract

Before utilising preclinical Position Emission Tomography (PET) systems for biological studies, evaluating their performance is important to better qualify the scanner’s applications. This study aims to assess the performance of the new extended field of view (FOV) nanoScan® PET/CT P123S system, developed for rodent total-body PET applications.

Scanner resolution, noise equivalent count rate (NECR), sensitivity and image quality were evaluated following NEMA NU-4 2008 protocols. Furthermore, a Derenzo phantom and linearity measurements were conducted. In vivo studies were subsequently carried out to evaluate system performance in biological applications.

The scanner spatial resolution according to the NEMA protocol was 1.4 mm using FBP reconstruction, while with iterative reconstruction it was under 0.7 mm. The NECR peak using a 250‒750 keV energy window was 1805.0 kcps at 93.7 MBq and 880.7 kcps at 88.4 MBq for the mouse-sized and rat-sized phantom respectively. The absolute sensitivity was 10.5%. The standard deviation of the uniform area of the image quality phantom was 1.8%, while the recovery coefficients varied between 0.23 and 1.00. The spill-over ratios were 0.04, and 0.04 in the water and air-filled chambers respectively. Quantitative bias was < 4% with a linear response up to 105 MBq. Total-body rat images were successfully acquired using the new system.

The new extended FOV PET system has improved sensitivity and count rate performance compared with previous systems. Its spatial resolution and quantitative accuracy are well-suited for preclinical PET applications. The extended FOV enables total-body imaging of both mice and rats.

The online version contains supplementary material available at 10.1186/s40658-025-00817-5.

## Linked entities

- **Species:** Mus musculus (taxon 10090)

## Full-text entities

- **Species:** Rattus norvegicus (brown rat, species) [taxon 10116], Mus musculus (house mouse, species) [taxon 10090], Rodentia (rodent, order) [taxon 9989]
- **Mutations:** P123S

## Figures

7 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12779859/full.md

---
Source: https://tomesphere.com/paper/PMC12779859