Comparison of the diagnostic performance of Enterosystem 18R and the BD Phoenix automated identification and susceptibility testing system for identifying gram-negative Enterobacteria: a preliminary study
Bwambale Jonani, Patricia Peace Nakitende, Daniella Nabakka, Herman Roman Bwire, Emmanuel Charles Kasule, John Bosco Mundaka

Abstract
Genes, proteins, chemicals, diseases, species, mutations and cell lines named across the full text — each resolved to its canonical identifier and authoritative record.
| Confidence threshold (%) | No. of isolates | % agreement | 95% CI (%) | Cohen’s Kappa (κ) |
|---|---|---|---|---|
| ≥50 | 11 | 45.5 | 16.7–76.6 | 0.32 |
| ≥70 | 8 | 62.5 | 24.5–91.5 | 0.48 |
| ≥85 | 7 | 71.4 | 29.0–96.3 | 0.58 |
| ≥90 | 6 | 66.7 | 22.3–95.7 | 0.48 |
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Taxonomy
TopicsBacterial Identification and Susceptibility Testing · Antibiotic Resistance in Bacteria · Vibrio bacteria research studies
LETTER
Resource-limited laboratories require affordable alternatives to automated bacterial identification systems, which cost approximately 35 per test in low-resource settings. We evaluated the Enterosystem 18R (which costs approximately $7 per test) (1), a semi-automated biochemical identification system, against the BD Phoenix automated identification system (Model M50, panel NMIC/ID-104) using 20 pre-characterized Gram-negative, oxidase-negative enterobacterial isolates from clinical urine specimens.
Twenty blinded isolates were obtained from Makerere University’s Microbiology Laboratory, originally characterized using standard biochemical protocols and confirmed by Bruker Biotyper MALDI-TOF (Version 4.1.100, Bruker Daltonics, 2019). Isolates were tested using Enterosystem 18R at Sebbi Hospital and BD Phoenix at the reference laboratory. Agreement was assessed at multiple confidence thresholds (≥50%, ≥70%, ≥85%, and ≥90%) for the Enterosystem 18R’s on-panel organisms only.
BD Phoenix identified all 20 isolates (100%), of which 11 were on-panel organisms for the Enterosystem 18R. When comparing these 11 on-panel organisms, the agreement between the systems was 45.5% (κ = 0.32) at ≥50% confidence. The agreement improved to 71.4% (κ = 0.58) at ≥85% confidence, but this restricted the analysis to only 7 of the 11 on-panel organisms (Table 1). Critical misidentifications occurred even with high confidence, including Klebsiella pneumoniae identified as Serratia plymuthica (98% confidence) and Acinetobacter baumannii as Yersinia pestis (88% confidence).
Difficulties in color interpretation in biochemical reactions significantly contributed to identification failures. Several reactions showed similar color transitions that were challenging to reliably differentiate, particularly the yellow-orange to red changes in the decarboxylation tests (wells 2–4, 7). These subjective color interpretations likely contributed to both identification failure and discordant results. Unlike automated optical reading, this semi-automated system relies on the subjective human interpretation of reaction color transitions, which can vary with operator experience and lighting conditions.
The 45% failure rate for on-panel organisms and poor diagnostic agreement suggests limited clinical utility despite cost advantages. Laboratories would require alternative identification methods for nearly half of the isolates, negating cost savings and perpetuating empirical antibiotic prescriptions without organism-specific guidance. The risk is that this would continue to perpetuate irrational antimicrobial prescriptions when reliable pathogen identification is impossible (2, 3).
This preliminary finding is limited by the small sample size (n = 20), single-site evaluation, and lack of a systematic inter-operator variability assessment. Larger multi-site validation studies with molecular adjudication of discordant results are needed to inform clinical implementation of the Enterosystem 18R microorganism identification system, especially in low-resource settings that are already constrained by a lack of microbiological diagnostic capability.
These preliminary findings indicate that Enterosystem 18R faces significant operational technical challenges that could limit its utility in resource-limited clinical settings. Future improvements to the system should prioritize minimizing the subjective reaction color interpretation by integrating an automated optical reader or alternative detection methods with distinct endpoints. Nevertheless, its affordability may still offer an interim solution to enhance microbial identification and support antimicrobial stewardship in low-resource settings.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1Liofilchem. 2025. Enterosystem 18R: biochemical Identification of Gram-negative, oxidase-negative Enterobacteria. Liofilchem. Available from: https://www.liofilchem.com/solutions/clinical/microbial-identification/id-systems/id-ast-systems-pure-colony/enterosystem-18r. Retrieved 26 Jun 2025.
- 2Wi TE, Ndowa FJ, Ferreyra C, Kelly-Cirino C, Taylor MM, Toskin I, Kiarie J, Santesso N, Unemo M. 2019. Diagnosing sexually transmitted infections in resource-constrained settings: challenges and ways forward. J Int AIDS Soc 22:e 25343. doi:10.1002/jia 2.2534331468679 PMC 6715950 · doi ↗ · pubmed ↗
- 3AMR Review. 2015. Rapid diagnostics: stopping unnecessary use of antibiotics. The review on antimicrobial resistance. https://amr-review.org/sites/default/files/Paper-Rapid-Diagnostics-Stopping-Unnecessary-Prescription-Low-Res.pdf.