Closed-Loop Automated Oxygen Control in Preterm Infants Receiving Non-Invasive Respiratory Support
Ourania Kaltsogianni, Theodore Dassios, Anne Greenough

TL;DR
Automated oxygen control systems help maintain proper oxygen levels in preterm infants using non-invasive ventilation, but long-term effects remain unclear.
Contribution
This review identifies that CLAC improves oxygen saturation target achievement but highlights the lack of long-term clinical outcome data.
Findings
CLAC systems increase the time preterm infants spend within target oxygen saturation ranges.
CLAC reduces time spent in extreme oxygenation levels compared to manual control.
Most studies on CLAC are short-term (≤24 hours) with no data on long-term clinical outcomes.
Abstract
What are the main findings? •Closed-loop automated oxygen control (CLAC) systems improve the achievement of oxygen saturation targets in preterm infants on non-invasive ventilation.•The evidence regarding prolonged use of CLAC is limited, and there are no reports on clinical outcomes. Closed-loop automated oxygen control (CLAC) systems improve the achievement of oxygen saturation targets in preterm infants on non-invasive ventilation. The evidence regarding prolonged use of CLAC is limited, and there are no reports on clinical outcomes. What are the implications of the main findings? •Future studies should explore the effect of the prolonged use of CLAC with different modes of non-invasive ventilation on oxygen saturation targeting and long-term outcomes. Future studies should explore the effect of the prolonged use of CLAC with different modes of non-invasive ventilation on oxygen…
Genes, proteins, chemicals, diseases, species, mutations and cell lines named across the full text — each resolved to its canonical identifier and authoritative record.
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Taxonomy
TopicsNeonatal Respiratory Health Research · Respiratory Support and Mechanisms · Retinopathy of Prematurity Studies
1. Introduction
Preterm infants frequently require supplemental oxygen, but, although its use can be life-saving, it can increase the risk of complications [1]. Hyperoxia increases oxidative stress [2] and increases the risks of bronchopulmonary dysplasia (BPD) and retinopathy of prematurity (ROP) [3,4]. Hypoxia increases morbidity and mortality [5,6].
Targeting oxygen therapy to maintain peripheral oxygen saturation levels (SpO_2_) within a predefined range can reduce the risk of complications. Compliance with achievement of oxygen saturation targets, however, has been shown to be rather poor and variable [7].
Closed-loop automated oxygen control (CLAC) systems may represent a solution to poor compliance and may, therefore, reduce complications. CLAC systems monitor oxygen saturation values in real time to calculate and make an adjustment to the inspired oxygen concentration (FiO_2_) without any human intervention [8]. Several algorithms exist to model and account for the relationship between SpO_2_ and FiO_2_ in infants with different severities of respiratory disease [9]. Studies have demonstrated that automated compared to manual oxygen control systems in preterm ventilated infants result in an improvement in the achievement of oxygen saturation targets, reduced time spent in hypoxia and hyperoxia and fewer manual adjustments to the inspired oxygen concentration [10,11,12,13,14,15,16,17,18,19,20,21,22,23,24]. In addition, CLAC has been associated with earlier weaning of the inspired oxygen [19,24]. In a randomised controlled trial (RCT) in preterm ventilated infants (n = 69), CLAC was associated with shorter durations of mechanical ventilation (MV) and supplemental oxygen and reductions in the incidence of BPD and the proportion of infants discharged on home oxygen [25].
Preterm infants are increasingly managed on non-invasive respiratory support with an aim to minimise lung injury and reduce respiratory morbidity [26]. Optimisation of supplemental oxygen treatment could further improve respiratory outcomes in this population. This narrative review aims to explore the efficacy of CLAC systems in preterm infants receiving non-invasive respiratory support and highlight areas for further research.
2. Methods
A literature search was conducted in PubMed for articles related to the use of CLAC in preterm infants. An advanced search was performed with a combination of search terms ((closed-loop) or (automat*) AND (oxygen) AND (neonat*). The reference list of relevant articles was also reviewed. We included published studies that compared closed-loop automated oxygen control to manual oxygen control in preterm infants (22 to 36 + 6 weeks of gestational age) on non-invasive respiratory support. Studies on CLAC that included only infants on invasive mechanical ventilation and studies comparing different automated oxygen controllers were excluded.
3. Results
Sixteen studies were identified that included preterm infants on non-invasive respiratory support: twelve randomised crossover studies, three RCTs and a matched-cohort study (Table 1). There were no studies involving term-born infants. A variety of algorithms, oxygen saturation targets, timing and durations of monitoring periods were used.
3.1. Current Evidence for the Use of Automated Oxygen Control in Preterm Infants on Non-Invasive Respiratory Support
3.1.1. CLAC with Non-Invasive Respiratory Support
Three randomised crossover studies compared CLAC with manual oxygen control in preterm infants with respiratory distress syndrome (RDS) who were receiving continuous positive airway pressure (CPAP) or high-flow nasal cannula (HFNC) oxygen [17,22,28]. The studies demonstrated an improvement in the achievement of oxygen saturation targets [17,22,28], reduced frequency and duration of hypoxemic [22,28] and hyperoxemic episodes and fewer manual adjustments to the FiO_2_ with automated oxygen control [17,22,28].
In a two-centre, randomised crossover study, Schwarz and co-workers compared two versions of a CLAC algorithm, which had different response times to SpO_2_ changes (CLAC_fast_ and CLAC_slow_) with manual oxygen control. The revised, faster algorithm for CLAC, which allowed more frequent FiO_2_ adjustments, improved SpO_2_ target achievement (68% versus 58%; p < 0.001), reduced the percentage of time spent below the targeted range (14% versus 22%; p < 0.001) and the number of manual FiO_2_ adjustments compared with manual control (p = 0.08). Moreover, there was a trend towards less time spent below oxygen saturation targets (SpO_2_ < 90%) with CLAC_fast_ compared with CLAC_slow_ (22% versus 17%; p = 0.16), suggesting that faster responsiveness may be more effective [27].
More recent data support that the efficacy of CLAC in infants receiving non-invasive respiratory support may differ over time. A single-centre RCT on the prolonged use (28 days) of automated oxygen control in extremely preterm infants found an increase in the time spent below the target range (SpO_2 _< 88% and SpO_2 _< 90%) in the first days of life that gradually declined. The opposite trend was observed during manual oxygen control [29].
3.1.2. High- or Low-Flow Nasal Cannula Oxygen
The beneficial effects of automated oxygen control on oxygen saturation targeting have also been observed in infants receiving high-flow nasal cannula (HFNC) oxygen treatment [24,30,31,32]. A RCT in extremely low birth weight infants on day eight to nine of life, showed that automated oxygen control over 12 h was associated with improved SpO_2_ target achievement (58% versus 33%; p < 0.01) and a lower time spent in hyperoxia (SpO_2_ > 95%; 26.5% versus 54.8%, p < 0.01) but an increase in the percentage of time spent with SpO_2_ < 85% (14% versus 11%; p = 0.05). The frequencies of prolonged hypoxic episodes and the incidence of more severe hypoxemia (SpO_2_: 70–75%) were reduced [30]. In another study, the mean FiO_2_ was higher during automated oxygen control (mean difference 0.019 (95% CI 0.006 to 0.03); p = 0.003), but there was no significant difference in the time spent in severe hyperoxia (SpO_2_ > 98%). This was attributed to staff training in how to avoid hypoxia and a potentially oversensitive response of the CLAC algorithm that led to overshoot and hyperoxic episodes [31]. A larger RCT that enrolled participants within 72 h of life and for the whole duration of HFNC treatment demonstrated a 26% increase in the percentage of time spent within oxygen saturation targets (p < 0.001) and reduced the time spent in extremes of SpO_2_ (<80% and >98%; p = 0.002 and <0.001, respectively) [32].
3.1.3. CLAC with Mechanical Ventilation or Continuous Positive Airway Pressure (CPAP)
Many studies compared CLAC with manual oxygen control in preterm infants receiving either invasive mechanical ventilation (MV) or nasal CPAP. The majority of them have a randomised crossover design [12,20,23,27,33], and another two studies were parallel arm randomised controlled trials [34,35]. Two multi-centre randomised crossover studies demonstrated that CLAC significantly improved compliance with achievement of oxygen saturation targets and the number of manual adjustments to the FiO_2_ compared with routine manual oxygen control [12,33]. These effects were present across different SpO_2_ target ranges [12,33]. Subgroup analyses in infants receiving invasive and non-invasive respiratory support showed comparable treatment effects of CLAC and reduced time spent in hypoxia (SpO_2 < 80%) or hyperoxia (SpO_2 > 98%) in the two groups [33]. Those studies, however, were of limited duration (24 h periods of automated and manual oxygen control) and tested different devices operating different algorithms [12,33].
Waitz et al. demonstrated that automated oxygen control improved oxygen saturation targeting (76.3% versus 69.1%, p < 0.01) among preterm infants receiving different modes of respiratory support (one ventilated; five on non-invasive positive pressure ventilation; nine on CPAP), but did not significantly alter the stability of cerebral tissue oxygenation over 24 h [20]. In addition, CLAC in very-low-birth-weight (VLBW) infants led to a 10% increase in the time spent within the intended oxygen saturation range (77.8% versus 68.5%, p < 0.001) and reduced incidence of prolonged hypoxemic episodes (SpO_2 _< 88%, > 180 s), but oxygen tissue saturation was not significantly affected [23].
An RCT in preterm infants on MV or nasal CPAP did not demonstrate any significant differences in the time spent in the target SpO_2_ range (90–95%) and the time spent in hyperoxia (SpO_2_ > 97% when FiO_2 > 0.21) between infants receiving automated or manual oxygen control [34]. CLAC use was associated with a significant reduction in the time spent in hypoxemia (SpO_2 < 80%; 0.1% versus 0.6%, p = 0.03) and the number of prolonged hypoxemic episodes (0.3 versus 2 per day, p = 0.03). The intervention, though, was applied only for 40% of the duration of respiratory support, as infants exited the study if they received support from another device. Therefore, the study results may not be generalisable to other controllers or for the whole duration of oxygen treatment [34]. In a larger RCT that enrolled infants with a mean gestation of 28.8 weeks on day one of life, CLAC increased the total time spent in normoxemia (p < 0.001), reduced the frequency of hypoxic and hyperoxic episodes and the number of manual FiO_2_ adjustments, but the monitoring duration was less than 24 h [35].
A recent matched cohort study demonstrated a time-varying effect of CLAC on oxygen saturation targeting [36]. Among 25 extremely preterm infants on different types of respiratory support, there were no significant differences in the achievement of oxygen saturation targets over a seven-week period. Automated oxygen control, however, was associated with a significant increase in the percentage of time spent within SpO_2_ target range during the first two weeks of life for those receiving supplemental oxygen (an increase of 9.9%, p = 0.01 in week one and 9.5%, p = 0.02 in week two), along with a reduced percentage of time spent in hyperoxia (SpO_2_ > 95%; −9.9% (95% CI [−19.2, −0.2]; p = 0.04)).
4. Discussion
This review highlights that automated oxygen control improves achievement of oxygen saturation targets in preterm infants receiving non-invasive respiratory support. A systematic review and meta-analysis included five studies in a subgroup of preterm infants on non-invasive ventilation (n = 169) and demonstrated a 15% increase in the time spent in the target SpO_2_ range with automated oxygen control [11]. In addition, a recent Cochrane review and meta-analysis showed that CLAC reduced the percentage of time spent below SpO_2_ target range (MD −5.95% (95% CI −7.98% to −3.92; five studies, 153 infants)) and in hyperoxia (MD −2.31% (95% CI −3.89% to −0.72; five studies, 153 infants; p < 0.001)) in infants receiving non-invasive respiratory support but the evidence was of moderate certainty [37]. Zhang et al. included in their meta-analysis studies in infants receiving MV or non-invasive respiratory support, but they did not report subgroup analysis based on the mode of ventilation [38]. Our narrative review focuses on the effectiveness of CLAC in preterm infants receiving different modes of non-invasive respiratory support. In addition, we include studies with prolonged use of the intervention [29,36], which suggest that the effect of CLAC on oxygen saturation targeting may differ over time, with potentially important implications for their use in the early neonatal period. Among extremely preterm non-ventilated infants, CLAC was associated with increased time spent below the target SpO_2_ range in the first few days of life compared with manual oxygen control [29]. In another study, automated oxygen control led to a significant improvement in the achievement of oxygen saturation targets in the first two weeks of life compared with manual control, but the difference was not significant for the total study duration [36].
There are, however, some limitations regarding the included studies. Most of them had a randomised crossover design, a few were prematurely stopped before reaching their predefined sample size [12,29,36,39] and study periods were less than or equal to 24 h [12,17,20,22,23,24,27,30,31,33,35,39]. Therefore, there was limited data available for analysis. Further, as infants were studied at different timepoints, the results may not reflect the effectiveness of CLAC for the whole duration of respiratory support. Indeed, two studies showed a time-varying effect of the intervention [29,36]. Schouten et al. found an increase in the time spent below the target SpO_2_ range with CLAC in the first few days of life that gradually declined [29], but the study was stopped prematurely. Dijkman and co-workers’ results supported the clinical importance of the use of CLAC in the early neonatal period to avoid extremes of oxygen saturations, but their study was limited by the risk of selection bias due to the matched cohort design and the exclusion of non-surviving participants [36]. Furthermore, many of the studies included infants receiving different modes of non-invasive respiratory support or mechanical ventilation and did not report results of subgroups. There were no studies including term-born infants, although some may require non-invasive respiratory support at birth, and previous studies demonstrated that use of CLAC in that population may have similar benefits as in the preterm population [40,41,42]. Furthermore, no data were found on any clinical outcomes for infants on non-invasive respiratory support who were receiving CLAC.
These limitations highlight the need for an adequately powered RCT of CLAC versus manual oxygen control in preterm non-ventilated infants and for the whole duration of non-invasive respiratory support. The trial should aim to determine the effectiveness of automated oxygen control with different modes of non-invasive respiratory support. In addition, staff training along with the use of CLAC can help reduce the incidence of hypoxia and hyperoxia and the risks related to them.
In conclusion, automated oxygen control improves achievement of oxygen saturation targets in preterm infants receiving non-invasive respiratory support by reducing the time spent outside the target SpO_2_ range. Future research should determine the effectiveness of the prolonged use of CLAC with different modes of non-invasive ventilation, with less heterogeneity of the studies in terms of timepoints and monitoring durations, and explore whether it improves long-term outcomes related to oxygen toxicity.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1Perrone S. Bracciali C. Di Virgilio N. Buonocore G. Oxygen Use in Neonatal Care: A Two-edged Sword Front. Pediatr.2016414310.3389/fped.2016.0014328119904 PMC 5220090 · doi ↗ · pubmed ↗
- 2Baba L. Mc Grath J.M. Oxygen free radicals: Effects in the newborn period Adv. Neonatal Care 2008825626410.1097/01.ANC.0000338015.25911.8a 18827514 · doi ↗ · pubmed ↗
- 3Supplemental Therapeutic Oxygen for Prethreshold Retinopathy Of Prematurity (STOP-ROP), a randomized, controlled trial. I: Primary outcomes Pediatrics 200010529531010.1542/peds.105.2.29510654946 · doi ↗ · pubmed ↗
- 4Askie L.M. Henderson-Smart D.J. Irwig L. Simpson J.M. Oxygen-saturation targets and outcomes in extremely preterm infants N. Engl. J. Med.200334995996710.1056/NEJ Moa 02308012954744 · doi ↗ · pubmed ↗
- 5The BOOST-II Australia United Kingdom Collaborative Groups Tarnow-Mordi W. Stenson B. Kirby A. Juszczak E. Donoghoe M. Deshpande S. Morley C. King A. Outcomes of Two Trials of Oxygen-Saturation Targets in Preterm Infants N. Engl. J. Med.201637474976010.1056/NEJ Moa 151421226863265 · doi ↗ · pubmed ↗
- 6Saugstad O.D. Aune D. In search of the optimal oxygen saturation for extremely low birth weight infants: A systematic review and meta-analysis Neonatology 20111001810.1159/00032200121150224 · doi ↗ · pubmed ↗
- 7Hagadorn J.I. Furey A.M. Nghiem T.H. Schmid C.H. Phelps D.L. Pillers D.A. Cole C.H. Group A.V.S. Achieved versus intended pulse oximeter saturation in infants born less than 28 weeks’ gestation: The AVI Ox study Pediatrics 20061181574158210.1542/peds.2005-041317015549 · doi ↗ · pubmed ↗
- 8Dani C. Automated control of inspired oxygen (Fi O 2) in preterm infants: Literature review Pediatr. Pulmonol.20195435836310.1002/ppul.2423830632296 · doi ↗ · pubmed ↗