# Sheathless delivery of a transfemoral pulsatile left ventricular assist device for high-risk percutaneous interventions: a case series

**Authors:** Edoardo Elia, Marcelo B Bastos, Demarchi Andrea, Giovanni La Malfa, Pistis Gianfranco, Giuseppe Patti, Gabriella Dallaglio, Gioel Gabrio Secco

PMC · DOI: 10.1093/ehjcr/ytaf483 · European Heart Journal. Case Reports · 2025-09-29

## TL;DR

A new method for inserting a heart support device without a large sheath was successfully tested in four high-risk cardiovascular procedures.

## Contribution

The study demonstrates a sheathless insertion technique for a left ventricular assist device during complex cardiovascular interventions.

## Key findings

- Sheathless insertion of the iVAC2L device was successfully performed in four cases without adverse events.
- The approach required an 11% smaller arteriotomy compared to traditional methods.
- The technique was applied in both coronary and structural heart interventions.

## Abstract

The use of mechanical circulatory support (MCS) can reduce adverse events in patients with severely impaired left ventricular (LV) function and precarious haemodynamics during complex cardiovascular procedures. Despite numerous benefits, MCS requires large-bore access, which increases the risks of vascular complications. The iVAC2L (PulseCath BV, Arnhem, The Netherlands) provides up to 2.0 L/min output using pulsatile flow and is an effective tool for LV unloading, but requires a large-bore sheath (18 Fr).

We describe four cases of sheathless insertion of iVAC2L during high-risk cardiovascular interventions. Two cases consisted in high-risk percutaneous coronary interventions, and two other cases were transcatheter edge-to-edge mitral valve repair. All cases were successful, and no adverse events could be observed.

The sheathless approach requires an 11% smaller arteriotomy. This case series aims to display how the iVAC2L catheter can be deployed without a sheath to minimize access size in a safe and feasible manner and proposes a structured approach to its deployment. When performed in selected patients, this strategy may reduce vascular complications.

Sheathless introduction of iVAC2L was feasible and safe safely deployed in selected cases of coronary and structural interventions. Larger studies are necessary to further define its impact on event rates.

## Full-text entities

- **Diseases:** impaired left ventricular (LV) function (MESH:D018487), vascular complications (MESH:D003925)
- **Chemicals:** iVAC2L (-)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12626125/full.md