# K-wires as a valid alternative to dual and single headless cannulated compression screw fixation of chauffeur fractures: a biomechanical comparison

**Authors:** Nicolas Catalin Ionut Ion, Bogdan-Axente Bocea, Ivan Zderic, Radu Sorin Fleaca, Mihai-Dan Roman, Cosmin-Ioan Mohor, R. Geoff Richards, Mark Lenz, Boyko Gueorguiev, Ludmil Drenchev, Tatjana Pastor

PMC · DOI: 10.1007/s00068-025-02971-y · European Journal of Trauma and Emergency Surgery · 2025-10-28

## TL;DR

This study compares different surgical methods for fixing Chauffeur fractures and finds that two K-wires work as well as two screws.

## Contribution

The study provides biomechanical evidence that two K-wires are a valid alternative to two headless cannulated compression screws for Chauffeur fractures.

## Key findings

- Two K-wires showed higher cycles to failure and failure load compared to one headless cannulated compression screw.
- Two headless cannulated compression screws had lower initial stiffness in flexion compared to K-wires.
- Two K-wires and two screws performed similarly in terms of failure resistance.

## Abstract

Chauffeur fractures represent a distinctive subset of distal radius fractures and typically arise from a high-energy axial load or a direct blow to the volar aspect of the wrist. In cases of significant displacement, joint incongruity or functional impairment, surgical intervention becomes crucial. However, up to date there is no consensus about the best osteosynthesis treatment option. Therefore, the aim of this study was to investigate the biomechanical competence of Chauffeur fractures fixed with either one headless cannulated compression screw, two headless cannulated compression screws, or two Kirschner (K-) wires.

Eighteen right synthetic radii with a simulated Chauffeur fracture were assigned to three groups (n = 6) for fixation with either one headless cannulated compression screw (3.0 mm x 36 mm; Group 1), two K-wires (1.8 mm; Group 2), or two headless cannulated compression screws (2.2 mm x 30 mm and 2.2 mm x 36 mm; Group 3). The specimens underwent non-destructive biomechanical testing in neutral position, flexion, and extension, followed by progressively increasing cyclic loading to failure in neutral position with monitoring via motion tracking.

Initial construct stiffness in flexion was significantly lower in Group 3 (202.4 ± 32.8 N/mm) versus both Group 1 (283.6 ± 31.3 N/mm) and Group 2 (275.3 ± 57.8 N/mm), p < 0.05. Cycles to failure and failure load in Group 2 (9214 ± 644; 510.7 ± 32.2 N) and Group 3 (8282 ± 973; 464.1 ± 48.7 N) were both significantly higher versus Group 1 (4678 ± 930; 283.9 ± 46.5 N), p < 0.05.

Whereas treatment of Chauffeur fractures with one cannulated compression headless screw is not recommended, using two K-wires is a valid alternative to the fixation of such fractures with two headless cannulated compression screws.

## Full-text entities

- **Diseases:** Chauffeur fracture (MESH:D050723), distal radius fractures (MESH:D000092503)

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12568812/full.md

## Figures

7 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12568812/full.md

## References

3 references — full list in the complete paper: https://tomesphere.com/paper/PMC12568812/full.md

---
Source: https://tomesphere.com/paper/PMC12568812