# Diagnostic Performance of CBCT in Detecting Different Types of Root Fractures with Various Intracanal Post Systems

**Authors:** Serhat Efeoglu, Ecem Ozgur, Aysenur Oncu, Ahmet Tohumcu, Rana Nalcaci, Berkan Celikten

PMC · DOI: 10.3390/tomography11100116 · Tomography · 2025-10-21

## TL;DR

This study compares two CBCT devices to see how well they detect different types of root fractures in teeth with various post systems.

## Contribution

The study evaluates the diagnostic accuracy of two CBCT devices under multiple imaging protocols for detecting root fractures with intracanal post systems.

## Key findings

- NewTom 7G showed almost perfect agreement with the gold standard in fracture detection.
- NewTom Go had reduced sensitivity for vertical fractures at lower dose settings.
- Both devices performed similarly in detecting horizontal and oblique fractures.

## Abstract

Objective: This study aimed to evaluate the diagnostic accuracy of two cone beam computed tomography (CBCT) devices using 18 imaging modalities in detecting root fractures—vertical, horizontal, and oblique—in teeth with intracanal post systems. Materials and methods: Ninety-six were extracted; single-rooted mandibular premolars were endodontically treated and restored with Bundle, Reforpost, or Splendor Single Adjustable posts. Controlled fractures of different types were induced using a universal testing machine. Each tooth was scanned with NewTom 7G and NewTom Go (Quantitative Radiology, Verona, Italy) under nine imaging protocols per device; varying in dose and voxel size, yielding 1728 CBCT images. Three observers (a professor of endodontics; a specialist; and a postgraduate student in endodontics) independently evaluated the images. Results: Observers demonstrated almost perfect agreement (κ ≥ 0.81) with the gold standard in fracture detection using NewTom 7G. No significant differences were found in sensitivity, specificity, or accuracy across voxel size and dose parameters for both devices in detecting fracture presence (p > 0.05). Similarly, both devices displayed comparable performance in identifying horizontal and oblique fractures (p > 0.05). However, in NewTom Go, regular and low doses with different voxel sizes showed reduced sensitivity and accuracy in detecting vertical fractures across all post systems (p ≤ 0.05). Conclusions: NewTom 7G, with its advanced detector system and smaller voxel sizes, provides superior diagnostic accuracy for root fractures. In contrast, NewTom Go displays reduced sensitivity for vertical fractures at lower settings. Clinical relevance: CBCT device selection and imaging protocols significantly affect the diagnosis of vertical root fractures.

## Full-text entities

- **Diseases:** fracture (MESH:D050723), Root Fractures (MESH:D011843), vertical fractures (MESH:D009759)

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12568167/full.md

## Figures

6 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12568167/full.md

## References

34 references — full list in the complete paper: https://tomesphere.com/paper/PMC12568167/full.md

---
Source: https://tomesphere.com/paper/PMC12568167