# Effect of Abutment Screw Design on Torque Loss Under Cyclic Fatigue Loading: A Comparison of TSIII and KSIII Implant Systems

**Authors:** Jung-Tae Lee, Jae-Chang Lee, Dong-Wook Han, Bongju Kim

PMC · DOI: 10.3390/bioengineering12101131 · Bioengineering · 2025-10-21

## TL;DR

This study compares two dental implant systems and finds that one (KSIII) is better at maintaining screw stability under stress, potentially reducing mechanical complications.

## Contribution

The study introduces a novel comparison of torque loss in two implant systems (TSIII and KSIII) under cyclic loading, revealing design-dependent biomechanical stability.

## Key findings

- KSIII showed significantly lower torque loss rates (below 25%) compared to TSIII (36.5% to 51.8%) across all torque levels.
- Torque loss in TSIII was 2.5–3.0 times higher than in KSIII, indicating superior stability in KSIII.
- A tightening torque of 35 Ncm provided the most stable preload in TSIII, but KSIII maintained consistent stability regardless of torque.

## Abstract

Background: Abutment screw loosening (ASL) is the most frequent mechanical complication in dentistry, leading to prosthetic instability and biological risks. Preload, generated during screw tightening, is critical for maintaining stability but is influenced by torque application, screw geometry, and cyclic loading. Methods: This in vitro study compared torque loss between two implant systems (Osstem TSIII and KSIII) with different abutment screw designs. Fifty implant–abutment assemblies (n = 5 per torque group) were tested under tightening torques of 20, 25, 30, 35, and 40 Ncm. Initial removal torque (T1) was measured 5 min after tightening, followed by cyclic loading (150 N, 14 Hz, 100,000 cycles). Post-fatigue removal torque (T2) was then recorded, and torque loss rate (%) was calculated. Independent t-tests and a one-way ANOVA were used for statistical analysis. Results: KSIII consistently exhibited higher T1 and T2 values than TSIII across all torque levels (p < 0.05). The torque loss rate for TSIII ranged from 36.5% (35 Ncm) to 51.8% (20 Ncm), showing a torque-dependent trend (p < 0.05). In contrast, KSIII maintained torque loss rates below 25% at all levels, with no significant differences between torque groups (p > 0.05). On average, torque loss in TSIII was approximately 2.5–3.0 times higher than in KSIII. Conclusions: The KSIII system demonstrated superior biomechanical stability, with significantly lower torque loss compared with TSIII, independent of torque level. Clinically, these findings suggest that the KSIII system may reduce the incidence of screw loosening and associated complications. A tightening torque of approximately 35 Ncm appeared to provide the most stable preload. Long-term in vivo studies are warranted to confirm these results under clinical conditions.

## Full-text entities

- **Diseases:** Fatigue (MESH:D005221), ASL (MESH:D011475)
- **Chemicals:** KSIII (-)

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12561121/full.md

## Figures

3 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12561121/full.md

## References

31 references — full list in the complete paper: https://tomesphere.com/paper/PMC12561121/full.md

---
Source: https://tomesphere.com/paper/PMC12561121