# Role of tube current variation and metal artifact reduction tool in CBCT for detection of missing MB2 canal

**Authors:** Caio Alencar-Palha, Marcyele Natane da Silva Morais, Fernanda Bulhões Fagundes, Lucas P. Lopes Rosado, Deborah Queiroz Freitas, Frederico Sampaio Neves

PMC · DOI: 10.1590/0103-644020256491 · 2025-10-24

## TL;DR

This study investigated how adjusting tube current and using a metal artifact reduction tool in CBCT affects the detection of a missing MB2 canal in molars.

## Contribution

The study provides evidence that tube current variation and MAR tool activation do not significantly impact MB2 canal detection in CBCT.

## Key findings

- AUC values ranged from 0.90 to 0.94 for MB2 canal detection.
- No significant differences were found in detection with varying tube current or MAR activation.
- A tube current of 4 mA without MAR activation is recommended for optimal dose-benefit ratio.

## Abstract

We aimed to evaluate the effect of varying the tube current and the Metal Artifact Reduction (MAR) tool on the detection of the missing second mesiobuccal (MB2) canal. Forty maxillary molars (20 with and 20 without MB2) were selected and confirmed by micro-CT. The teeth underwent endodontic preparation, with the canals filled using gutta-percha and a silver-palladium metal post. CBCT scans were acquired using the OP300 Maxio system (voxel size: 0.085 mm, FOV: 50 x 50 mm) at 4 mA, 6.3 mA, and 12.5 mA, with and without MAR activation. The areas under the receiver operating characteristic curves (AUCs), sensitivities, and specificities were calculated for five examiners and compared using two-way ANOVA (p < 0.05). Intra- and interexaminer agreements were assessed using the weighted Kappa test. The AUC values ranged from 0.90 to 0.94, sensitivity from 0.75 to 0.78, and specificity from 0.89 to 0.95. Intra- and interexaminer indices ranged from moderate to substancial. Based on the tube current settings or MAR activation, there were no significant differences in the MB2 canal detection. Therefore, varying the tube current and MAR tool activation did not affect MB2 canal detection, with values close to 4 mA without MAR activation recommended in this scenario to reflect a better dose-benefit ratio.

## Linked entities

- **Chemicals:** silver-palladium (PubChem CID 51341926)

## Full-text entities

- **Chemicals:** silver (MESH:D012834), palladium (MESH:D010165)