# Systematically lower aEEG amplitude values in neurologically healthy children using an automated algorithm compared to a semi-manual method

**Authors:** Sandra Greve, Pia Brensing, Luisa Paul, Christian Dohna-Schwake, Julia Wichlacz, Ursula Felderhoff-Mueser, Nora Bruns

PMC · DOI: 10.3389/fneur.2025.1562580 · 2025-10-08

## TL;DR

This study compares automated and semi-manual methods for measuring aEEG amplitudes in children and finds that automated methods yield systematically lower values.

## Contribution

The study introduces a systematic comparison of automated versus semi-manual aEEG amplitude measurements in pediatric patients.

## Key findings

- Semi-manual amplitude measurements were consistently higher than automated assessments.
- Mean differences in upper amplitudes ranged from 23.7 to 29.3 μV across channels.
- Age-related trends in aEEG amplitudes were observed despite differing absolute values between methods.

## Abstract

The objective of this study was to compare different modes of amplitude-integrated EEG (aEEG) assessment (semi-manual vs. automated) in children.

A total of 450 unremarkable pediatric EEGs from children aged 6 months to 17.9 years were converted into aEEGs and the medians and means of the upper and lower amplitudes (C3–P3, C4–P4, C3–C4, P3–P4, Fp1–Fp2) were determined. The agreement of the semi-manual and automated measurements was assessed via the Pearson correlation coefficients (PCC) and Bland-Altman plots. Mean differences between the methods and age-specific percentiles (5th−95th) were calculated.

Semi-manually measured amplitudes were systematically greater than automated assessments. Mean differences of the means ranged between 23.7 and 29.3 μV for the upper and between 2.4 and 4.4 μV for the lower amplitudes depending on the channel. The PCC ranged between 0.68 and 0.92 for the upper and lower amplitudes of the mean depending on the channel. Age-specific percentiles showed different absolute values but similar trends.

AEEG amplitude values systematically differ between semi-manual and automated assessment. Age-related trends are evident despite differences in the absolute values. Reference values for different measurement techniques are needed for pediatric aEEG.

## Full-text entities

- **Diseases:** epilepsy (MESH:D004827), neurologic disease (MESH:D020271), Seizures (MESH:D012640)
- **Chemicals:** C-P (-)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Figures

3 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12547116/full.md

---
Source: https://tomesphere.com/paper/PMC12547116