# Volume targeted mask ventilation during simulated neonatal resuscitation – A randomized crossover manikin study

**Authors:** Chelsea Morin, Kashmala Yousafzai, Brenda Hiu Yan Law, Georg M. Schmölzer

PMC · DOI: 10.1016/j.resplu.2025.101109 · Resuscitation Plus · 2025-09-23

## TL;DR

A study compared different ventilation devices during simulated neonatal resuscitation and found that the NextStepTM device reduced mask leak significantly.

## Contribution

The study introduces a novel ventilation device (NextStepTM) and evaluates its performance in reducing mask leak during neonatal resuscitation.

## Key findings

- The NextStepTM device significantly reduced mask leak compared to other devices like the Draeger Ventilator and T-piece.
- The NextStepTM delivered consistent tidal volumes with less variation and lower peak inflation pressure.
- Using the NextStepTM with a two-hand hold improved mask leak reduction in a neonatal manikin model.

## Abstract

To compare mask positive pressure ventilation (PPV) provided by pressure guided devices (i.e., T-Piece) with or without a respiratory function monitor (RFM) with ventilator-based volume-targeted ventilation (VTV) using a VN500 Draeger ventilator or the NextStepTM, a novel ventilation device designed for the delivery room.

Prospective, randomized, crossover, simulation study. Following orientation to ventilation devices, participants were randomized to order of four ventilation devices (NextStepTM, VN500 Draeger ventilator, T-piece PPV with RFM visible, and T-piece PPV with RFM masked) and order of four simulation scenarios. The study was performed in a neonatal resuscitation room within a level 3 neonatal intensive care unit. Participants were trained neonatal resuscitation providers or instructors with experience as team leader. Semi-automated, ventilator-based volume-targeted mask PPV (VTV-PPV) (NextStepTM or Draeger Ventilator) was compared to manual PPV via a T-piece device (RFM either visible or masked). Primary outcome was reduction in mask leak with the NextStepTM compared to the other devices.

Thirty-two healthcare professionals [25 (78.1 %) were female and 7 (21.9 %) were male] participated. The median (interquartile range) mask leak was significantly lower with VTV-PPV with NextStepTM [6 (1–12) %] compared to the Draeger Ventilator [24 (25–38)%, p = 0.01], T-Piece with RFM [18 (9–33) %, p = 0.0088], and T-Piece without RFM [32 (12–57)%, p = >0.0001]. The median (IQR) delivered tidal volume was not different between groups, although the NextStepTM had less tidal volume variation compared to all other groups and peak inflation pressure was significantly lower with VTV-PPV with NextStepTM compared to all other groups.

In a neonatal manikin model, VTV-PPV with the NextStepTM using a two-hand hold reduced mask leak compared to the T-piece without RFM guidance.

## Full-text entities

- **Diseases:** leak (MESH:D019559)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12538474/full.md

## Figures

2 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12538474/full.md

## References

38 references — full list in the complete paper: https://tomesphere.com/paper/PMC12538474/full.md

---
Source: https://tomesphere.com/paper/PMC12538474