# Agreement of Corneal Measurements With Swept-Source Optical Coherence Tomography (Anterion) and Scheimpflug (Sirius) Device in Eyes With Dermatochalasis

**Authors:** Hatice Kubra Sonmez, Metin Unlu, Cem Evereklioglu, Fatih Horozoglu, Osman Ahmet Polat, Hatice Arda

PMC · DOI: 10.7759/cureus.89802 · Cureus · 2025-08-11

## TL;DR

This study compares two devices for measuring corneal parameters in patients with dermatochalasis and finds they provide similar, reliable results.

## Contribution

The study evaluates the agreement between Anterion and Sirius devices in corneal measurements for patients with dermatochalasis.

## Key findings

- Anterion and Sirius showed excellent agreement in most corneal measurements despite minor statistical differences.
- Posterior astigmatism showed moderate agreement but strong correlation between the two devices.
- The differences observed were clinically insignificant, allowing for interchangeable use of the devices.

## Abstract

Introduction

This study aimed to evaluate the agreement between two advanced anterior segment imaging devices, Anterion (Heidelberg Engineering GmbH, Heidelberg, Germany) and Sirius (CSO, Florence, Italy), in measuring keratometric and pachymetric parameters in patients with moderate to severe dermatochalasis.

Methods

Keratometric and pachymetric measurements were obtained using Anterion and Sirius devices in patients with moderate to severe dermatochalasis. The agreement between the two devices was assessed by comparing anterior and posterior corneal keratometry and corneal thickness. Statistical analyses were performed to determine the significance of differences and correlations between the measurements.

Results

Statistically significant differences were observed between the two devices in anterior and posterior corneal keratometry, except for posterior astigmatism and corneal thickness. However, these differences were clinically insignificant(low refractive differences). The study found excellent agreement between Anterion and Sirius for most parameters, with strong correlations and minimal disparities in measurements. Posterior astigmatism showed moderate agreement but still exhibited a strong correlation between the devices.

Conclusion

Both Anterion and Sirius provide reliable keratometric and pachymetric measurements in patients with dermatochalasis, allowing for interchangeable use. Accurate preoperative evaluations are essential for cataract or refractive surgery, as dermatochalasis may affect corneal measurements.

## Full-text entities

- **Diseases:** astigmatism (MESH:D001251), cataract (MESH:D002386)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12340465/full.md

## Figures

2 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12340465/full.md

## References

15 references — full list in the complete paper: https://tomesphere.com/paper/PMC12340465/full.md

---
Source: https://tomesphere.com/paper/PMC12340465