Autorefraction Pitfalls in Extended Depth-of-Focus Pseudophakia: A Case Report
Andreas F Borkenstein, Eva-Maria Borkenstein

TL;DR
This case report highlights how autorefraction can mislead after EDOF IOL implantation, causing incorrect prescriptions and patient discomfort.
Contribution
The paper introduces a novel case series showing how EDOF IOLs can cause misleading autorefraction results and emphasizes the need for subjective refraction.
Findings
Three patients with EDOF IOLs showed myopia on autorefraction but had excellent uncorrected vision.
Incorrect spectacle prescriptions led to asthenopic symptoms, resolved upon discontinuation.
Subjective refraction and functional assessment are critical after EDOF IOL implantation.
Abstract
Extended depth-of-focus (EDOF) intraocular lenses (IOLs) are increasingly used to provide a broad range of vision after cataract surgery. However, their unique optical design can produce misleading results when postoperative refraction is assessed using automated methods. We report three pseudophakic patients implanted bilaterally with Tecnis PureSee IOLs, all of whom demonstrated excellent uncorrected visual acuity (1.0) but appeared myopic on autorefractor testing. Based on these values, opticians prescribed minus-powered spectacles, including night-driving glasses and progressive lenses. All three patients developed asthenopic symptoms such as headaches, nausea, diplopia, or dizziness. Symptoms resolved rapidly upon discontinuation of spectacles. However, diagnosis was delayed by 2.5 to 4 weeks. These cases illustrate the risk of overcorrection due to objective refraction artifacts…
Genes, proteins, chemicals, diseases, species, mutations and cell lines named across the full text — each resolved to its canonical identifier and authoritative record.
| Case | Age (years)/Sex | IOL type/Target refraction | UCDVA (binocular, decimal) | UCNVA (binocular, decimal at distance in cm) | Autorefraction | Prescribed spectacle type | Symptoms |
| 1 | 64/F | Tecnis PureSee/+-0.00 OD and -0.25 OS | 1.0 | 0.7 at 50 cm | -1.00/-0.25 at 100° OD and -1.25/-0.25 at 90° OS | Sunglasses + driving glasses (night/bad weather) | Headache, nausea, ocular burning |
| 2 | 78/M | Tecnis PureSee/+-0.00 OU | 1.0 | 0.8 at 55 cm | -1.00/-0.50 at 20° OD and -1.00/0.00 OS | Driving glasses + theatre use | Dizziness, discomfort, visual confusion |
| 3 | 67/F | Tecnis PureSee/-0.50 OU | 1.0 | 0.8 at 40cm | -1.50/0.00 OD and -1.25/0.00 OS | Driving glasses + progressive lens for near/household tasks | Headache, instability, diplopia |
| Case | Eye | Keratometry K1 (D) | Keratometry K2 (D) | Anterior chamber depth (mm) | Axial length (mm) | Implanted IOL power (D) | Target refraction (D) |
| Case 1 | OD | 42.71 | 43.11 | 3.2 | 23.84 | 20.5 | 0 |
| Case 1 | OS | 42.61 | 42.89 | 3.6 | 24.73 | 18.0 | -0.25 |
| Case 2 | OD | 41.98 | 42.43 | 3.1 | 24.65 | 18.5 | 0 |
| Case 2 | OS | 42.43 | 42.76 | 3.9 | 25.12 | 17.0 | 0 |
| Case 3 | OD | 42.98 | 43.32 | 4.0 | 26.8 | 12.0 | -0.50 |
| Case 3 | OS | 42.78 | 43.01 | 4.2 | 25.9 | 15.0 | -0.50 |
| Mechanism | Category | Explanation | Clinical impact |
| IR focal shift | Measurement-related | Infrared light focuses behind the visible light plane due to its longer wavelength | Devices misjudge the focal point, creating an apparent myopic error |
| Wavefront-based interpretation | Device limitation | Autorefractors interpret the return wavefront shape; steeper wavefronts are read as increased negative power | Systematic overestimation of myopia, especially in pseudophakic eyes |
| Chromatic dispersion | Material-related | Hydrophobic acrylic IOLs with low Abbe numbers cause wavelength-dependent focal shifts | Amplifies the discrepancy between the IR and visible spectrum, increasing false myopia |
| Central optical modulation | Optical | EDOF IOLs reshape the central optical zone to extend the depth of focus | Alters the refractive profile, often misinterpreted by autorefractors |
| Depth of focus tolerance | Optical | EDOF lenses provide broad defocus tolerance, masking small residual errors | Patients maintain good vision despite objective error, leading to unnecessary prescriptions |
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Taxonomy
TopicsIntraocular Surgery and Lenses · Retinal and Macular Surgery · Ophthalmology and Visual Impairment Studies
Introduction
In recent years, intraocular lens (IOL) technologies have evolved beyond the restoration of distance visual acuity after cataract surgery. Modern designs increasingly target functional vision at various distances, enhanced contrast sensitivity, and reduced reliance on spectacles. This shift has led to the development and widespread use of so-called “premium” IOLs. These lenses address diverse visual needs and have redefined postoperative expectations for both patients and clinicians.
Extended depth-of-focus (EDOF) IOLs are increasingly used in cataract surgery to improve functional vision across distance and intermediate ranges while minimizing dysphotopsia. Their non-diffractive optical design aims to preserve contrast sensitivity and reduce photic phenomena compared to traditional multifocal lenses [1-3].
Postoperative refraction is routinely assessed using automated refractometers in both surgical and community optometric settings. While efficient and widely available, autorefraction may produce misleading results in eyes implanted with modern IOL optics, particularly EDOF designs [4-6].
A characteristic phenomenon is the appearance of a myopic shift, despite excellent uncorrected distance acuity, resulting in inappropriate spectacle prescriptions when interpreted without awareness of lens-specific behavior [6,7].
This case report presents three patients implanted with Tecnis PureSee EDOF lenses who experienced asthenopic symptoms due to overcorrection based on autorefraction alone. These cases highlight the importance of subjective verification, interdisciplinary communication, and awareness of the limitations of wavefront-based measurement techniques in pseudophakic eyes.
Case presentation
Case 1
A 64-year-old woman with bilateral cataracta senilis underwent an uneventful bilateral cataract extraction with implantation of Tecnis PureSee IOLs, aiming for emmetropia in the right eye (0.00 sph) and slight myopia (-0.25 sph) in the left eye. At the one-week postoperative follow-up, uncorrected distance visual acuity (UCDVA) was 0.9 (decimal) in both eyes and 1.0 (decimal) binocular, and binocular near visual acuity (UCNVA) reached 0.7 (decimal) using a reading distance of 50 cm, without need for correction. The patient reported high visual satisfaction at this stage. However, postoperative autorefraction measurement showed -1.00 D/-0.25 D cyl at 100° OD and -1.25 D/-0.25 D cyl at 90° OS. Based on these readings, the optician recommended a distance prescription for UV-blocking sunglasses and clear driving glasses intended for nighttime driving or bad weather conditions. After wearing those glasses, the patient developed continuous headaches, nausea, and ocular burning, which persisted over several hours. Symptoms resolved rapidly after discontinuation of spectacle use. The functional mismatch between objective measurements and the patient’s subjective visual experience was identified only after this episode.
Case 2
A 78-year-old man underwent uncomplicated bilateral cataract surgery with implantation of Tecnis PureSee IOLs, targeting emmetropia in both eyes (0.00 sph OU). At the one-week postoperative follow-up, UCDVA was 1.0 (decimal) in each eye and 1.0 (decimal) binocularly. UCNVA reached 0.8 (decimal) at a reading distance of 55 cm, and no correction was required. The patient reported high subjective satisfaction and no visual complaints. At a follow-up visit with his optician, postoperative autorefraction measured -1.00 D/-0.50 D cyl at 20° OD and -1.00 D/0.00 D cyl OS. Based on these findings, the optician recommended prescription spectacles for driving and attending theater performances. Shortly after beginning to use the glasses, the patient developed significant dizziness, ocular discomfort, and visual confusion, particularly under dim lighting conditions such as the theater. These symptoms subsided promptly following discontinuation of the spectacles. The mismatch between objective measurements and the patient’s actual visual experience led to a delayed recognition of the problem. The patient expressed frustration regarding unnecessary spectacle costs and multiple follow-up visits, despite an otherwise successful surgical outcome.
Case 3
A 67-year-old woman with bilateral cataracts and moderate pre-existing myopia (-4.50 D sph) underwent uneventful cataract surgery with implantation of Tecnis PureSee IOLs. The surgical target was a slight residual myopia of -0.50 D in both eyes to support near function. At the one-week postoperative follow-up, UCDVA was 0.9 (decimal) in both eyes and 1.0 (decimal) binocularly. UCNVA reached 0.8 (decimal) at a reading distance of 40 cm, and no optical correction was required at this time. The patient reported high visual satisfaction in daily activities. Two weeks later, at a routine visit with her optician, postoperative autorefraction showed -1.50 D in the right eye and -1.25 D in the left eye. Based on these measurements, she was prescribed two pairs of spectacles: one for distance vision (e.g., nighttime driving) and one progressive addition lens for near work and household tasks. Soon after initiating spectacle use, the patient developed persistent frontal headaches, instability while walking, and transient diplopia during gaze shifts. These symptoms improved markedly after she stopped wearing the glasses. The underlying issue, overcorrection based on autorefractor readings in an otherwise functionally emmetropic patient, was identified three weeks later. She expressed clear dissatisfaction due to the unnecessary financial burden of approximately €900 spent on the spectacle corrections (Table 1).
The data reinforce our interpretation that the observed myopic shift on autorefraction was artifactual rather than anatomical or surgical (Table 2).
Discussion
The presented cases underscore the limitations of objective refraction when applied to eyes implanted with advanced (“premium”) IOLs. All three patients demonstrated excellent UCDVA yet received myopic spectacle prescriptions based solely on autorefractor data. This mismatch between functional vision and automated measurements reflects the interaction between modern IOL optics and infrared-based metrology.
According to the European Society of Cataract and Refractive Surgeons (ESCRS), premium IOLs can be grouped by optical mechanism and intended visual performance into several categories [1]. Enhanced monofocal IOLs maintain a monofocal profile but incorporate asphericity or refractive transitions to improve intermediate vision (e.g., Tecnis Eyhance ICB00) [2]. EDOF IOLs provide a continuous range from far to intermediate, using diffractive or non-diffractive wavefront-shaping elements (e.g., Tecnis PureSee DEN00V and Tecnis Odyssey DRN00V) [3,4]. Multifocal or hybrid IOLs (e.g., Tecnis Synergy DFR00V) extend focus further into the near range by combining multiple focal planes [5,6]. While non-diffractive EDOF lenses preserve image quality and reduce glare, they can interfere with standard measurement techniques. In particular, a falsely myopic result is often observed on autorefraction, even in the presence of excellent visual function [6-8]. This phenomenon is likely multifactorial (Table 3).
First, EDOF lenses modulate the central wavefront to stretch the depth of focus, which can alter how automated devices detect the return image. Second, autorefractors typically use infrared (IR) light (800-900 nm) to assess refractive error. Although IR light focuses posterior to the retina (suggesting a hyperopic displacement), the returning wavefront is steeper and interpreted by the device as increased negative power. Autorefractors do not assess the focal point directly but infer refractive status from wavefront curvature. This wavefront-based analysis leads to a systematic overestimation of myopia in pseudophakic eyes with EDOF optics [7-9].
Moreover, the chromatic dispersion of IOL materials, especially hydrophobic acrylics with low Abbe numbers, enhances wavelength-dependent focal differences, contributing further to IR-visible mismatch [8].
Clinically, this artifact may go unnoticed unless the patient reports symptoms. All three of our patients were functionally emmetropic, yet developed headaches, dizziness, or diplopia after using minus lenses prescribed based on autorefraction alone (“overcorrection”). In all cases, symptoms resolved after stopping spectacle use. To confirm that the observed myopic readings were not attributable to biometric error or incorrect IOL selection, we presented key preoperative biometric data in Table 2. This includes axial length, corneal power (K1/K2), anterior chamber depth, and implanted IOL power with target refraction. The calculations were performed using the ESCRS calculation program, and at least two formulas (e.g., Barrett and Kane) were used as cross-checks. Postoperative subjective refraction further supports that no significant residual refractive error was present.
The biometric data and the good uncorrected visual acuity support the assumption that true refractive error was minimal and not due to surgical miscalculation, consistent with previous findings on biometry accuracy in EDOF implantation [10].
From a systems perspective, these cases expose a communication gap between cataract surgeons and community opticians. Without access to IOL model information or awareness of its optical profile, opticians may unknowingly overcorrect eyes that functionally do not require it. Implementing IOL identification cards, digital records, or structured communication protocols could help reduce such errors. Patients often maintain good visual performance across a range of defocus values, even with small residual refractive errors, as shown in prior studies on EDOF lens performance [11-13]. Clinicians should interpret postoperative refraction in the context of subjective patient feedback and neuroadaptation. In patients with EDOF lenses and satisfactory visual function, delaying spectacle prescription is often the safest course, especially when automated refraction suggests minor myopia without clinical symptoms [9,14].
Overall, successful postoperative care of premium IOL patients depends not only on surgical precision but also on interdisciplinary coordination and refractive awareness across all care providers involved.
Conclusions
EDOF IOLs provide excellent functional visual outcomes, particularly for distance and intermediate vision. However, their unique optical design may lead to misleading postoperative measurements when standard autorefractors are used. In all cases presented, patients demonstrated functionally emmetropic outcomes with excellent uncorrected visual acuity, yet were mistakenly overcorrected based on device-generated myopic readings. These cases underscore a critical gap between objective refraction and real-world visual function in pseudophakic eyes with EDOF optics. Importantly, symptoms such as asthenopia and visual discomfort resolved only after discontinuation of the unnecessary spectacles. To optimize outcomes and avoid patient dissatisfaction, a refined postoperative approach is essential. This includes delaying spectacle prescriptions until subjective visual stability is confirmed, prioritizing patient-reported visual quality over automated metrics, and fostering consistent communication between surgeons, refractionists, and optical professionals. Understanding the optical behavior of EDOF lenses and tailoring refractive care accordingly should be standard practice in the management of premium IOL patients.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1European Society of Cataract and Refractive Surgeons. ESCRS guideline for cataract surgery: IOL classification framework 6 2025 European Society of Cataract and Refractive Surgeons.(2023 02023 https://www.escrs.org/escrs-guideline-for-cataract-surgery/
- 2Visual outcome, optical quality, and patient satisfaction with a new monofocal IOL, enhanced for intermediate vision: preliminary results J Cataract Refract Surg Mencucci R Favuzza E Venturi D Vignapiano R Favuzza E 378385462020 https://doi.org/10.1097/j.jcrs.00000000000000613205021810.1097/j.jcrs.0000000000000061 · doi ↗ · pubmed ↗
- 3Johnson & Johnson Vision. Tecnis Pure See™ IOL: technical whitepaper Tecnis Pure See™ IOL: Technical 6 2025 Johnson & Johnson Vision.(2023 2023 https://jnjvisionpro.eu/products/tecnis-puresee-iol
- 4Johnson & Johnson Vision. Tecnis Odyssey™ Extended Range of Vision IOL - Product brochure 6 2025 Johnson & Johnson Vision.(2024 2024 https://www.jnjvisionpro.com/en-us/products/tecnis-odyssey/
- 5Evaluation of visual and subjective outcomes with mix-and-match of three one-piece Tecnis multifocal IO Ls of varying add powers Clin Ophthalmol Dell S 290329111420203306127610.2147/OPTH.S 266120 PMC 7532892 · doi ↗ · pubmed ↗
- 6Johnson & Johnson Vision. Tecnis Synergy™ IOL - clinical overview Tecnis Synergy™ IOL - Clinical 6 2025 Johnson & Johnson Vision.(2023 2023 https://jnjvisionpro.eu/synergy
- 7Functional assessment of a new extended depth-of-focus intraocular lens Eye (Lond) Savini G Balducci N Carbonara C 4044103320193026698510.1038/s 41433-018-0221-1PMC 6460699 · doi ↗ · pubmed ↗
- 8Optical bench evaluation of the effect of pupil size in new generation monofocal intraocular lenses BMC Ophthalmol Alarcon A Canovas C Koopman B Pande MV Koch DD Piers P 1122320233694162410.1186/s 12886-023-02839-y PMC 10026412 · doi ↗ · pubmed ↗