# Performance validation of Abbott Alinity i chemiluminescence analyzer for five thyroid function tests

**Authors:** Hongyu Zhang, Baixiu Wu, Liuhua Ke, Zheng Peng

PMC · DOI: 10.1016/j.plabm.2025.e00487 · Practical Laboratory Medicine · 2025-06-25

## TL;DR

This study confirms the Abbott Alinity i analyzer performs reliably for five thyroid tests, meeting quality standards for accuracy and precision.

## Contribution

The study provides empirical validation of the Abbott Alinity i for thyroid function tests according to CLSI guidelines.

## Key findings

- Precision and accuracy met quality targets for all five thyroid tests.
- Linearity was confirmed for T3, T4, and TSH within allowable deviation limits.
- Sample carryover effect was below 1%, meeting performance requirements.

## Abstract

To verify and evaluate the performance of the Abbott Alinity i chemiluminescence analyzer for five thyroid function tests.

Referring to the relevant documents of the Clinical and Laboratory Standards Institute (CLSI) and related literature, the precision, accuracy, linear range, reference interval, and sample carryover effect of Abbott Alinity i immunoassay system for measuring FT3, FT4, T3, T4, and TSH were verified and analyzed.

Precision: Repeatability ranged from 1.23 % to 6.11 %, and intermediate precision ranged from 1.84 % to 7.33 %, both meeting the quality targets (≤6.25 % and ≤8.33 %, respectively). Accuracy: The deviation between the mean value and the target value was less than 12.5 %, indicating good agreement. Linearity: For T3, T4, and TSH, the 95 % confidence intervals of the deviation from linearity (difference between measured and expected value) were entirely within the allowable deviation limits (ADL). For FT3 and FT4, manufacturer guidelines preclude dilution; thus, linearity verification was omitted. Reference interval: All test values of 20 healthy individuals were within the reference intervals provided by the manufacturer. Sample carryover effect: The sample carryover effect ranged from −0.4 % to 0.17 %, which met the requirement of less than 1 %.

The Abbott Alinity i analyzer demonstrated acceptable performance in precision, accuracy, linearity (for T3, T4, and TSH), reference interval, and carry-over effect for the five thyroid function tests, meeting manufacturer specifications.

## Full-text entities

- **Diseases:** thyroid function (MESH:D013966)
- **Chemicals:** T4 (MESH:D013974), T3 (MESH:D014284), FT3 (-)

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12269562/full.md