# Optimization of secondary flow path clearance in centrifugal blood pump: a combined numerical and experimental study

**Authors:** Chenying Zhu, Ping Ye, Zhaohua Chang

PMC · DOI: 10.3389/fphys.2025.1595588 · Frontiers in Physiology · 2025-06-27

## TL;DR

This study optimizes the design of centrifugal blood pumps to improve performance and reduce blood damage using simulations and experiments.

## Contribution

The study introduces an optimal clearance combination in centrifugal blood pumps to balance leakage and turbulence.

## Key findings

- Numerical predictions matched experimental data with a maximum pressure deviation of 9%.
- Changes in clearance affected hemolysis performance more than head or efficiency.
- Asymmetric clearance may balance leakage control and turbulence suppression.

## Abstract

Blood pumps as mechanical circulatory support (MCS) devices are widely used for patients with ventricular heart failure, and improving blood compatibility remains a key focus for researchers.

The effects of secondary flow path clearance on the performance and hemocompatibility of a centrifugal blood pump are investigated in this study. A blood pump of the specific design was developed for modeling and computer simulation, and 25 models were generated by varying its internal and outer clearance between 600 and 1,000 μm. This study systematically investigates the hydrodynamic and hemodynamic performance of designed blood pumps through integrated computational fluid dynamics simulations and in vitro hydraulic experiments.

Key findings reveal a strong agreement between numerical predictions and experimental data, showing a maximum pressure deviation of 9% for the baseline pump configuration. Furthermore, the effect of the change in clearance on the head and efficiency is less than ±3%. The change in the secondary flow path significantly impacts hemolysis performance.

An optimal clearance combination exists in the designed model, and for a specific flow field structure, asymmetric clearance may be an effective means to balance leakage control and turbulence suppression. Although the findings are specific to the pumps examined in this study, they provide valuable insights into the optimal design of blood pumps.

## Full-text entities

- **Diseases:** hemolysis (MESH:D006461), ventricular heart failure (MESH:D006333)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12245888/full.md

## Figures

9 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12245888/full.md

## References

30 references — full list in the complete paper: https://tomesphere.com/paper/PMC12245888/full.md

---