# Transcatheter Repair of a Sinus Venosus Defect With a Modified Harmony Transcatheter Pulmonary Valve Prosthesis

**Authors:** Allison L. Tsao, Stephanie Colello, Matthew J. Gillespie

PMC · DOI: 10.1016/j.jaccas.2025.103294 · JACC Case Reports · 2025-02-26

## TL;DR

A modified heart valve was used to successfully repair a large heart defect in an 83-year-old patient who couldn't undergo traditional stent-based treatment.

## Contribution

An innovative use of a modified transcatheter pulmonary valve for transcatheter repair of a large sinus venosus atrial septal defect.

## Key findings

- The modified HTPV22 frame successfully sealed the SVASD without obstructing pulmonary venous return.
- The procedure was effective in a patient with cardiac dimensions unsuitable for standard stents.
- One-month follow-up showed complete sealing and stable anchoring of the device.

## Abstract

Transcatheter repair of sinus venosus atrial septal defect (SVASD) with partial anomalous pulmonary venous return (PAPVR) has become an acceptable surgical alternative for appropriately selected patients. Dilated cardiac anatomy may limit candidacy because of size limitations in balloon-expandable covered stents (BECSs). An 83-year-old man with SVASD with PAPVR presented with decompensated right heart failure secondary to a long-standing left-to-right shunt. Cardiac dimensions precluded the use of a BECS, and we modified a Harmony transcatheter pulmonary valve 22 mm (HTPV22, Medtronic) by removing the leaflets to use the self-expanding frame to cover the SVASD. Following release, the HTPV22 frame was anchored with additional BECSs to secure the frame within the superior vena cava. At 1-month follow-up, there was complete sealing of the SVASD and unobstructed pulmonary venous return to the left atrium. This report describes an innovative modification of the HTPV22 to allow for successful transcatheter repair of a large SVASD.

## Linked entities

- **Diseases:** atrial septal defect (MONDO:0006664)

## Full-text entities

- **Diseases:** SVASD (MESH:C548009), PAPVR (MESH:D012587), right heart failure (MESH:D006333)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12245424/full.md

## Figures

7 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12245424/full.md

## References

9 references — full list in the complete paper: https://tomesphere.com/paper/PMC12245424/full.md

---
Source: https://tomesphere.com/paper/PMC12245424