# A study of the clinical usefulness of the eMEMO®

**Authors:** Hinako Nakayama, Nobuhisa Watanabe, Misa Tsukamoto, Miyu Mori, Eri Nakakubo, Shogo Watanabe, Manabu Taniguchi

PMC · DOI: 10.1002/joa3.70137 · Journal of Arrhythmia · 2025-07-09

## TL;DR

This study shows that the waterproof eMEMO® device is effective for detecting heart rhythm issues, even during activities like bathing.

## Contribution

The study demonstrates the eMEMO®'s ability to detect arrhythmias during bathing, which is not possible with traditional 24-hour Holter monitoring.

## Key findings

- The eMEMO® detected arrhythmias in 21.3% of cases over extended monitoring periods.
- Arrhythmic events occurred during bathing in 52 patients, highlighting the device's utility in such scenarios.
- The device showed minimal skin irritation and low noise contamination during use.

## Abstract

Taking advantage of the characteristics of the novel compact external loop recorder (ELR; eMEMO®), which is waterproof and has an automatic arrhythmia detection and monitoring function for up to 14 days, we investigated the usefulness of ELR measurement, especially during bathing.

We included 294 patients (102 males and 192 females) who presented with palpitations as their main complaint from April 2021 to September 2022, with a mean age of 57.0 ± 17.4 years. The average wearing period of the ELR was 4.5 ± 0.9 days.

ELR for longer than 24 h led to the first arrhythmia detection in 170 (21.3%) of the 799 cases of all detected arrhythmias, including 11 with premature atrial contraction, 50 with premature ventricular contraction, 8 with atrial fibrillation, 32 with supraventricular tachycardia, and 21 with sick sinus syndrome. They had low noise contamination during measurement. Arrhythmic events during bathing were observed in 52 patients, effectively detecting arrhythmia during bathing, which was difficult with the usual 24‐h Holter electrocardiography (ECG). In addition, the itching and rash of the skin at the electrode attachment over a period of up to 5 days were very mild, indicating the comfort of the waterproof patch‐type electrode.

The patch‐type ELR (eMEMO®) with waterproof design is capable of continuous recording regardless of restrictions such as bathing or exercise and is useful for detecting stable arrhythmias in patients with palpitations.

Taking advantage of the characteristics of the novel compact external loop recorder (ELR; eMEMO®), we investigated the usefulness of ELR measurement. As a result, the patch‐type ELR, which is waterproof and can continuously record regardless of restrictions such as bathing or exercise, is useful for detecting stable arrhythmias in patients.

## Linked entities

- **Diseases:** atrial fibrillation (MONDO:0004981), sick sinus syndrome (MONDO:0001823)

## Full-text entities

- **Diseases:** arrhythmia (MESH:D001145), premature atrial contraction (MESH:D018880), rash (MESH:D005076), supraventricular tachycardia (MESH:D013617), premature ventricular contraction (MESH:D018879), Arrhythmic (OMIM:212500), sick sinus syndrome (MESH:D012804), itching (MESH:D011537), atrial fibrillation (MESH:D001281), palpitations (MESH:D006331)
- **Chemicals:** eMEMO (-)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12238776/full.md

## Figures

5 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12238776/full.md

## References

25 references — full list in the complete paper: https://tomesphere.com/paper/PMC12238776/full.md

---
Source: https://tomesphere.com/paper/PMC12238776