# Validation of the Somnolyzer 24×7 automatic scoring system in children with suspected obstructive sleep apnea

**Authors:** Ignacio Boira, Violeta Esteban, José Norberto Sancho-Chust, Esther Pastor, Paula Fernández-Martínez, Anastasiya Torba, Eusebi Chiner

PMC · DOI: 10.3389/fmed.2025.1617530 · Frontiers in Medicine · 2025-06-18

## TL;DR

This study validates the Somnolyzer 24×7 system for automatically scoring sleep data in children suspected of having sleep apnea, showing strong agreement with expert manual scoring.

## Contribution

The study provides empirical validation of the Somnolyzer 24×7 system's accuracy in children with suspected obstructive sleep apnea.

## Key findings

- The Somnolyzer 24×7 system showed high agreement with manual scoring of the respiratory disturbance index (RDI).
- Correlation between manual and automatic scoring was consistent across different sleep stages.

## Abstract

Manual scoring of polysomnography data is a laborious and complex process. Automatic scoring by current computer algorithms shows high agreement with manual scoring. The primary objective of this study was to measure the overall validity of the Somnolyzer 24×7 automatic polysomnography scoring system in children.

We conducted a single-center, prospective, observational study in children undergoing diagnostic polysomnography for suspected obstructive sleep apnea (OSA) from December 2023 to December 2024. We included children aged three to 15 years with suspected obstructive sleep apnea (OSA). Each polysomnogram was scored manually by three experts and automatically by the Somnolyzer 24×7 system.

Our analysis included 75 children (60% girls), of whom 9% did not have OSA, 20% had mild OSA, 31% moderate OSA, and 40% severe OSA. There was a high level of agreement between manual and automatic scoring of the respiratory disturbance index (RDI). The mean correlation (Pearson correlation coefficient) of RDI scored by the three experts was 0.93 (95% confidence interval [CI] 0.92–0.95), similar to the correlation between manual and automatic scoring (0.92, 95% CI 0.90–0.94). The correlation between the different manual scorings and between manual and automatic scoring was maintained in the different sleep stages (N1: 0.93 vs. 0.90, N2: 0.76 vs. 0.73, N3: 0.72 vs. 0.76, REM: 0.86 vs. 0.82).

The Somnolyzer 24×7 automatic scoring system shows strong correlation with manual scoring in respiratory events and sleep architecture. Our results suggest this system could be used for polysomnography scoring in children.

## Linked entities

- **Diseases:** obstructive sleep apnea (MONDO:0007147)

## Full-text entities

- **Diseases:** OSA (MESH:D020181), respiratory disturbance (MESH:D012131)

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12213436/full.md

## Figures

5 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12213436/full.md

## References

34 references — full list in the complete paper: https://tomesphere.com/paper/PMC12213436/full.md

---
Source: https://tomesphere.com/paper/PMC12213436