# Real‐World Impact of Steerable Sheaths for Paroxysmal Atrial Fibrillation Catheter Ablation: The INSIGHT Study

**Authors:** Yasuo Okumura, Tsunesuke Kono, Akira Mizukami, Osamu Inaba, Yuji Wakamatsu, Daisuke Yamagishi, Satoshi Nakamura, Takumi Arashiro, Akira Sato, Yi Wang, Atsushi Kobori

PMC · DOI: 10.1111/jce.16658 · Journal of Cardiovascular Electrophysiology · 2025-04-01

## TL;DR

The VIZIGO sheath reduces fluoroscopy time and dose during heart procedures for atrial fibrillation without affecting safety or effectiveness.

## Contribution

Real-world evidence showing the VIZIGO sheath improves procedural efficiency in paroxysmal atrial fibrillation ablation.

## Key findings

- VIZIGO sheath significantly reduced fluoroscopy time and dose compared to non-steerable sheaths.
- Procedural times for PVI isolation were shorter with the VIZIGO sheath.
- Acute and long-term effectiveness and safety were comparable between the two sheath types.

## Abstract

In atrial fibrillation (AF) ablation procedures, the CARTO VIZIGO Bidirectional Guiding Sheath has previously shown promising results in reducing fluoroscopy times without compromising clinical effectiveness or safety compared with non‐steerable sheaths.

This non‐randomized, multicenter, retrospective cohort study (INSIGHT) aimed to determine the real‐world impact of the VIZIGO sheath on procedural efficiency, clinical effectiveness, and safety in paroxysmal AF (PAF) catheter ablation. Consecutive adults who underwent de novo radiofrequency (RF) ablation for PAF with a non‐VIZIGO sheath (January 2019–July 2021) or VIZIGO sheath (January 2020–July 2021) were included. Procedural characteristics, primary adverse events (PAEs), and 12‐month effectiveness (freedom from repeat ablation or recurrent atrial arrhythmia) were evaluated.

Of 199 patients included (mean age, 69.7 years; 59.3% male), 97 had ablation with a VIZIGO sheath and 102 with a non‐VIZIGO sheath. Significantly shorter mean times were achieved in the VIZIGO vs. non‐VIZIGO group for time to left PVI (18.1 ± 7.5 vs. 19.9 ± 5.6 min, p = 0.046), right PVI (16.5 ± 6.1 vs. 23.1 ± 9.9 min, p < 0.001), total PVI (34.6 ± 9.7 vs. 42.9 ± 11.4 min, p = 0.002), and fluoroscopy time (7.3 ± 10.4 vs. 18.3 ± 13.3 min, p < 0.001). Mean fluoroscopy dose was significantly lower (45.9 ± 112.0 vs. 139.5 ± 251.5 mGy, p < 0.001) with VIZIGO vs. non‐VIZIGO sheaths. Catheter stability was comparable between groups. PAE rates were similar in the VIZIGO (3.1%) and non‐VIZIGO (4.9%) groups. Freedom from repeat ablation and recurrent atrial arrhythmia at 12 months were also similar in the two groups.

These real‐world data demonstrate that use of the VIZIGO sheath in PAF ablation procedures allows for significantly lower fluoroscopy time and dose with significantly shorter PVI isolation time, without compromising acute and long‐term effectiveness or safety.

Reductions in fluoroscopy time and dose and shorter PVI isolation times with the VIZIGO vs. non‐VIZIGO sheath: results of the real‐world, multicenter, INSIGHT cohort study.

## Linked entities

- **Diseases:** atrial fibrillation (MONDO:0004981)

## Full-text entities

- **Diseases:** atrial arrhythmia (MESH:D001145), AF (MESH:D001281)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12160683/full.md

## Figures

4 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12160683/full.md

## References

19 references — full list in the complete paper: https://tomesphere.com/paper/PMC12160683/full.md

---
Source: https://tomesphere.com/paper/PMC12160683