# A goat experimental epistaxis model: Hemostatic effect of stop nosebleeds device

**Authors:** Amani Abu-Shaheen, Shroaq Saleh Aljanobui, Falah Hassan Almohanna, Mohsen Ayyash, Sumayyia Marar, Goran Matic, Mohammed Hazazi, Juneil Batalla, Maaweya Awadalla, Muaawia A. Hamza, Mehmet Baysal, Mehmet Baysal, Mehmet Baysal

PMC · DOI: 10.1371/journal.pone.0324727 · PLOS One · 2025-06-10

## TL;DR

A new device called 'Stop Nosebleeds' was tested on goats and showed faster stopping of nosebleeds compared to manual compression.

## Contribution

The study introduces a novel silicone-based device for epistaxis treatment and evaluates its hemostatic efficacy in a goat model.

## Key findings

- The device reduced bleeding duration to 45.5 seconds compared to 206.0 seconds with manual compression.
- It was effective in both heparinized and non-heparinized goats, stopping bleeding in 52.0 and 39.0 seconds, respectively.
- The study highlights the need for further human trials to confirm clinical applicability.

## Abstract

To stop epistaxis, a Saudi medical invention called “Stop Nosebleeds” with a patent was created. It is a silicone-based, adjustable device that is applied externally to the bridge of the nose and speeds up clotting by decreasing the temperature and the compressing effect of the device on the nose This study aims to examine the efficacy of the device in attaining hemostasis in goats with anterior nasal bleeding. An animal experimental nasal-bleeding model was conducted on ten goats in collaboration between King Fahad Medical City and King Faisal Specialist Hospital and Research Center, Riyadh, Saudi Arabia. The left nostril was used as interventional wounds and the right nostril was used as control. Control wounds were treated with manual compression for 20 minutes. Interventional wounds were treated with the Stop Nosebleeds device. The second group of animals was heparinized and 2 wounds were created in each animal, one of which was being treated with the Stop Nosebleeds device, while the other was being treated with manual compression. The mean bleeding duration in the experimental group (both heparinized and non-heparinized goats) was 45.5 seconds (SD = 8.2), compared to 206.0 seconds (SD = 75.7) in the control group (U = 0.00, p < 0.001). The device stopped bleeding in 52.0 seconds and 39.0 seconds in the heparinized and non-heparinized experimental groups. While, the manual compression of epistaxis stopped bleeding in 252.5 seconds (SD = 84.7) and 159.2 seconds (SD = 15.7) in the heparinized and non-heparinized control groups, respectively. The device shows promise for clinical application, however, further research with larger sample sizes and human clinical trials are needed to validate its efficacy and safety. Additionally, the use of a goat model presents limitations in directly translating these results to human clinical practice, and these differences should be careful.

## Full-text entities

- **Diseases:** bleeding (MESH:D006470), wounds (MESH:D014947), Nosebleeds (MESH:D004844)
- **Chemicals:** silicone (MESH:D012828)
- **Species:** Capra hircus (domestic goat, species) [taxon 9925], Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12151359/full.md

## Figures

5 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12151359/full.md

## References

28 references — full list in the complete paper: https://tomesphere.com/paper/PMC12151359/full.md

---
Source: https://tomesphere.com/paper/PMC12151359