# Translation of Animal Study to Human: In Silico Based Development of Implantable Pulmonary Artery Pressure Sensor

**Authors:** Leonid Goubergrits, Pavlo Yevtushenko, Adriano Schlief, Jan Romberg, Titus Kuehne, Andreas Arndt, Jan Bruening

PMC · DOI: 10.1002/cnm.70050 · International Journal for Numerical Methods in Biomedical Engineering · 2025-06-02

## TL;DR

Researchers used computer models to check if a heart pressure sensor from animal tests would work safely in humans.

## Contribution

A parallel in silico human study was proposed to confirm the transferability of animal findings to human conditions.

## Key findings

- Human pulmonary arteries have larger left and right diameters than pigs, but similar main artery diameters.
- Hemodynamic changes from sensor implantation in humans did not increase thrombus risk.
- Findings from animal studies were confirmed in human in silico models despite anatomical differences.

## Abstract

Implantable pulmonary artery pressure sensors (PAPS) might impose a flow‐induced risk of thrombus formation in the pulmonary artery (PA). To assess this risk, an in silico study‐enhanced animal study with 20 sensors implanted in 10 pigs had previously been conducted. In the in silico study, PAPS were virtually implanted mimicking real implantations, based upon data acquired by CT. This animal in silico study investigated changes in hemodynamics caused by PAPS using image‐based computational fluid dynamics (CFD). However, porcine and human PA differ significantly in geometry and hemodynamics. To investigate the transferability of animal in silico study findings toward human conditions, we propose a parallel in silico human study. Based on a similarity analysis (L1 norm for 8 geometric features) human PA geometries with the least difference to 10 porcine PA were selected. PAPS were virtually implanted in human PA as close as possible, mimicking the implantation configuration of the animal study. Finally, a numerical flow analysis of the hemodynamic changes due to PAPS implantation was done. Comparing human and porcine PA, we found significantly larger left and right PA diameters in humans, whereas no differences were found for main PA diameters and bifurcation angle. Comparing hemodynamic boundary conditions, we found a significantly smaller heart rate and a significantly higher peak systolic main PA flow rate in humans, whereas no significant differences for cardiac output were found. The human in silico PAPS study found no relevant changes in hemodynamics increasing the risk of thrombus formation after sensor implantation. This is also valid for PAPS that were non‐optimally implanted. Thus, despite differences between species, findings of the in silico animal study were confirmed by the human in silico study.

The CFD study investigated the impact of the pressure sensor implanted in the porcine or human pulmonary artery (PA) on hemodynamics, aiming to prove the transferability of animal trial results to human conditions. We found that despite differences between human and porcine PA regarding morphometry and flow conditions, the results of the animal study are also valid for human conditions.

## Linked entities

- **Species:** Sus scrofa (taxon 9823), Homo sapiens (taxon 9606)

## Full-text entities

- **Diseases:** thrombus (MESH:D013927)
- **Species:** Homo sapiens (human, species) [taxon 9606], Sus scrofa (pig, species) [taxon 9823]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12128572/full.md

## Figures

6 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12128572/full.md

## References

45 references — full list in the complete paper: https://tomesphere.com/paper/PMC12128572/full.md

---
Source: https://tomesphere.com/paper/PMC12128572