# A ceramic liner fracture during total hip arthroplasty: a case report

**Authors:** Huihui Zhou, Yan Wang

PMC · DOI: 10.1093/jscr/rjaf346 · Journal of Surgical Case Reports · 2025-05-30

## TL;DR

This case report describes a rare intraoperative ceramic liner fracture during a total hip arthroplasty procedure.

## Contribution

The report highlights a rare intraoperative complication of ceramic liner fracture during THA.

## Key findings

- Ceramic liner fracture occurred during insertion due to nonconcentric reduction.
- The case involved a woman with bilateral avascular necrosis and collapse.
- The complication expands understanding of rare THA risks.

## Abstract

The ceramic-on-ceramic bearing has been a reliable bearing option in total hip arthroplasty (THA) for both elderly and younger patients with a long life expectancy. However, ceramic bearings are at risk of fracture. This case describes a complication during THA involving the use of a cementless press-fit acetabular shell and ceramic liner for a woman with bilateral avascular necrosis and collapse. Ceramic liner fracture occurred during its insertion owing to nonconcentric reduction. Our case report helps to expand the understanding of this rare complication of THA: ceramic liner fracture can occur intraoperatively.

## Linked entities

- **Diseases:** avascular necrosis (MONDO:0018373)

## Full-text entities

- **Diseases:** collapse (MESH:D001261), Ceramic liner fracture (MESH:D050723), avascular necrosis (MESH:D010020)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12124477/full.md

## Figures

4 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12124477/full.md

## References

15 references — full list in the complete paper: https://tomesphere.com/paper/PMC12124477/full.md

---
Source: https://tomesphere.com/paper/PMC12124477