Comparison of Fat Suppression Sequences T2 Weighted Two-Point Dixon and Short-Tau Inversion Recovery in Magnetic Resonance Imaging of Lumbar Spine: An Observational Study
Umesh Khanal, Ramswarth Sah, Sushil Kumar Mahato, Shailendra Katwal, Ghanshyam Gurung

TL;DR
This study compares two MRI fat suppression techniques, Dixon and STIR, for lumbar spine imaging, finding Dixon provides better contrast and less noise.
Contribution
The study empirically compares the performance of Dixon and STIR sequences in lumbar spine MRI using Contrast Ratio and Contrast-to-Noise Ratio metrics.
Findings
Dixon sequence showed higher contrast ratio (0.44) and contrast-to-noise ratio (11.95) compared to STIR.
STIR had lower contrast metrics with a contrast ratio of 0.31 and contrast-to-noise ratio of 8.74.
Dixon sequence also demonstrated lower background noise than STIR.
Abstract
Magnetic Resonance Imaging is a common diagnostic tool used to evaluate various clinical conditions. Different fat suppression techniques such as Short Tau Inversion Recovery and Dixon are employed to enhance diagnostic accuracy. The choice of fat suppression sequence varies based on availability and performance. This study aimed to compare Contrast Ratio and Contrast-to-Noise Ratio of Short Tau Inversion Recovery and Dixon technique. This observational cross-section study was performed in the Department of Radiology from 8 September 2023 to 7 September 2024 after the approval by the Institutional Review Committee (Reference number: 151/080/081(6-11)E2). Using a complete census method during the study period, MRI-lumbar spines of 384 adult patients were included in the study. Signal intensity values of the lesion, adjacent normal tissue, and noise were recorded for calculation of…
Genes, proteins, chemicals, diseases, species, mutations and cell lines named across the full text — each resolved to its canonical identifier and authoritative record.
|
Parameters |
DIXON |
STIR |
|---|---|---|
|
Repetition Time (ms) |
3970 |
5530 |
|
Echo Time (ms) |
75 |
70 |
|
Echo Train Length |
15 |
20 |
|
Number of acquisitions |
1 |
1 |
|
Concatenations |
1 |
1 |
|
No. of slices |
15 |
15 |
|
Slice thickness(mm) |
4 |
4 |
|
Phase encoding direction |
Head to feet |
Head to feet |
|
Acquisition time |
2min 20 sec |
2 min 30 sec |
|
Phase oversampling (%) |
70 |
70 |
|
Feild of View |
300 X 300 |
300 X 300 |
|
Inversion time (ms) |
NA |
160 |
|
Sex |
Frequency (%) |
18-28 years |
28-38 years |
38-48 years |
48-58 years |
58-68 years |
68-78 years |
≥78 years |
|---|---|---|---|---|---|---|---|---|
|
Male |
191 (49.74) |
26 (13.60) |
39 (20.40) |
31 (16.20) |
35 (18.30) |
48 (25.10) |
10 (5.20) |
2 (1.00) |
|
Female |
193 (50.26) |
17 (8.80) |
33 (17.10) |
57 (29.50) |
42 (21.80) |
26 (13.50) |
12 (6.20) |
6 (3.10) |
|
STIR sequence |
Dixon sequence | ||||
|---|---|---|---|---|---|
|
SInormal |
SIlesion |
SInoise |
SInormal |
SIlesion |
SInoise |
|
94.80 (76.55-117.50) |
85.30 (62.07-120.77) |
3.20 (2.30-4.70) |
73.65 (56.70-100.60) |
65.90 (43.60-103.07) |
2.50 (1.70-3.80) |
|
SInormal = Signal Intensity on normal Tissue, SIlesion = Signal Intensity on Lesion, SInoise = Signal Intensity on Image background | |||||
|
CRDixon |
CRstir |
CNRDixon |
CNRstir |
Background NoiseDixon |
Background NoiseSTiR |
|---|---|---|---|---|---|
|
0.44 (0.19-0.74) |
0.31 (0.14-0.50) |
11.95 (5.38-23.18) |
8.74 (3.82-8.73) |
2.50 (1.70-3.80) |
3.20 (2.30-4.70) |
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Taxonomy
TopicsMedical Imaging and Analysis · Advanced MRI Techniques and Applications · Ultrasound and Hyperthermia Applications
INTRODUCTION
Fat suppression is a Magnetic Resonance Imaging (MRI) technique in which signal from adipose tissue is suppressed to better visualize contrast uptake by body's tissues, reduce chemical-shift artifact, characterize certain lesions and improve to increase the diagnostic confidence in lumbar spine diseases.^1,^ ^2^ Due to short relaxation times, fat has high signal that helps to characterize lesions but also causes artifacts such as ghosting, chemical shift and mask subtle contrast difference in non-fatty tissue by filling the dynamic range of receiver mostly with fat signal and hide contrast enhancing tumor by surrounding it.^3^
Fat suppression can be achieved in different ways. But, due to empirical and practical knowledge gap, incomplete fat suppression and anatomical deterioration are common in fat suppressed images.^4-7^ Thus, the aim of our study was to compare the effectiveness of Fat Suppression Sequences: T2-weighted Two-point Dixon and Short Tau Inversion Recovery (STIR) in MRI of lumbar Spine images in terms of Contrast Ration (CR) and Contrast to Noise Ratio (CNR).
METHODS
This observational cross-section study was conducted among 384 patients of age group 18-84 years visiting the Department of Radiology and Imaging, Tribhuvan University Teaching Hospital, Nepal for MRI of lumbar spine. A total population sampling was done which included images of all patients between September 08, 2023 and September 07. The study included all patients with lower back pain or radiculopathy necessitating lumbar spine MRI examinations and referred for lumbar spine MRI examinations in the Department of Radiology and Imaging excluding infants, children, and individuals with post-instrumentation implants.
Before commencing the study, ethical approval was obtained from the Institutional Review Committee of Institute of Medicine, Tribhuvan University (Reference number: 151/080/081(6-11)E2). Informed consent was obtained from the patients or the legal guardian before enrollment in the study. To minimize sedation time, the extra fat suppression technique (Dixon) was not taken in case of pediatric patients.
Data were acquired using the 1.5T Magnetom Amira Siemens MRI scanner. All patients underwent thorough screening by departmental guidelines to identify presence of any ferromagnetic materials and patients were positioned in the 1.5T Magnetom Amira Siemens MRI scanner. The routine department imaging sequences including Short Tau Inversion Recovery (STIR) was followed by an additional T2 Dixon sequence in sagittal plane using predefined parameters (Table 1).
After completion of scanning for all sequences, sagittal images of T1 weighted, T2 weighted, STIR and T2 Dixon so obtained were evaluated. Any abnormality and signal changes in lumbar spine were first observed and measurement was taken. The symmetrical (same size) of region of interest (ROI) was drawn in normal region, on the lesion (Modic changes) noted on body of lumbar spine and also on the background to measure signal intensity (SI) which was represented by mean/SD on lumbar spine images of STIR and T2 Dixon sequences respectively. Finally, contrast ratio (CR) and contrast to noise ratio (CNR) were calculated using the following formulae: CR = (S_Ilesion_ - S_Inormal_)/S_Inormal_ and CNR = (SI_lesion_ - S_Inormal_)/S_Inoise_ Where, CR is contrast ratio, CNR is contrast-to-noise ratio, SIlesion is signal intensity from lesion, SInormal is signal intensity from normal tissue and SInoise is signal intensity from background noise.^8^
A predesigned pro-forma was used for recording socio-demographic variables, SI, CR and CNR. Data were entered in Microsoft Excel and then exported to and analysed using Statistical Package for the Social Sciences (SPSS), version 27 (IBM Corp., Armonk, N.Y., USA) for descriptive analysis.
RESULTS
A total of 384 patients were included in the study. Age of the patient ranged from 18 years to 84 years with median age 46 (IQR: 35-60) years. Among the total 384 patients, 191 (49.74%) were males and 193 (50.26%) were female (Table 2).
The median SI value in normal tissue and in lesion were 94.80 (IQR: 76.55-117.5) and 85.30 (IQR: 62.07-120.77) respectively in the image of STIR sequence while they were found 73.65 (IQR: 56.70-100.6) in normal tissue and 65.90 (IQR: 43.60-103.07) in the lesion in the image of Dixon sequence. The median noise value in the image of STIR sequence was 3.20 (IQR: 2.30-4.70) while it was 2.50 (IQR: 1.70-3.80) in case of image of Dixon sequence (Table 3).
The Median values of CR for images obtained by STIR and Dixon sequences were 0.31 (IQR: 0.14-0.50) and 0.44 (IQR: 0.19-0.74) respectively. Similarly, the median value of CNR for the images obtained by STIR and Dixon sequences were found to be 8.74 (IQR: 3.82-8.73) and 11.95 (IQR: 5.38-23.18) respectively. The median value for background noise was 2.50 (IQR: 1.70-3.80) for images obtained by Dixon sequence while that for images obtained by STIR sequence, 3.20 (IQR: 2.30-4.70) A significant differences in CR, CNR and background noise of the images obtained by STIR and Dixon sequences (p-value <0.001 for CR, CNR and background noise and z-value 9.68 for CR, 10.14 for CNR and 14.29 for background noise) (Table 4).
DISCUSSION
Magnetic Resonance Imaging provides standardized, non-invasive assessments of spinal pathologies and thus is one of the most important imaging modality regarding the diagnosis and management of the spinal pathologies.^1, 9^ Fat suppression technique is crucial to better visualize uptake of contrast material by body's tissues, reduce chemical-shift artifact, characterize certain lesions (modic changes, degenerative disc disease, marrow edema) and improve to increase the diagnostic confidence in lumbar spine diseases.^1,^ ^2^ Understanding the different fat-suppression options allows radiologists to adopt the most appropriate technique for their clinical practice.^10^
Our study aimed to compare the degree of fat suppression in terms of CR, CNR and background noise in the images produced by Dixon sequence and that produced by STIR sequence. For quantitative evaluation of fat suppression and lesion conspicuity, the calculated median CR-value for images obtained by STIR and Dixon sequences were found 0.31 (IQR: 0.14-0.50) and 0.44 (IQR: 0.19-0.74) respectively and the median CNR calculated for the images obtained by STIR and Dixon sequences were found 8.74 (IQR: 3.82-8.73) and 11.95 (IQR: 5.38-23.18) respectively. There was a difference between the CR values of images obtained by STIR sequence and Dixon sequence along with the CNR values of images obtained by STIR sequence and Dixon sequence.
In our study, CR and CNR were found to be higher and background noise lower for Dixon compared to Short Tau Inversion Recovery.
Wohlgemuth et al. conducted a comparative study measuring CR and CNR in 27 patients and found that multipoint Dixon technique shows superior SNR and CNR than STIR in the delineation of traumatic fractures.^11^ Similarly, Lee S et al. conducted a similar comparative study in 79 patients and found T2 Dixon as superior to Spectral Attenuated Inversion Recovery (SPAIR) in quantitative parameters (better CR and CNR, p-value <0.01) which also support our study results.^1^ In spite of small sample size in the above two studies, they found similar result to our study findings which highlight the superiority of Dixon sequence as fat suppression technique.
Likewise, Zadig P et al. in 2023 conducted a study entitled Pediatric whole-body magnetic resonance imaging: comparison of STIR and T2 Dixon sequences in the detection and grading of high signal bone marrow changes and found that 90% of high signal bone marrow areas identified on a 1.5-T whole-body MRI were seen on both STIR and water-only T2W Dixon, while 5% were seen on STIR only and 5% were seen on T2W Dixon only. This result may be due to difference in inclusion criteria.^3^
Low RN et al. conducted a comparative study to evaluate a prototype fast spin-echo (FSE) tripleecho Dixon (FTED) technique for T2-weighted spine imaging with and without fat suppression compared to conventional T2-weighted fast recovery (FR) FSE and short-tau inversion recovery (STIR) imaging and found the FTED water images demonstrated less motion, better anatomic sharpness and less CSF, however, there was no difference in fat suppression and chemical shift artifact.^12^
Ma J et al. conducted a comparative study and found mean scores 3.2 and 2.1 for the uniformity of fat suppression, and 3.0 and 2.0 for the lesion conspicuity for the fast Three Point Dixon and the chemical shift selective (CHESS) fat suppression techniques, respectively. In spite of limited sample size 27, this study's result also showed fast three Point Dixon technique statistically superior to the CHESS technique (P < 0.0005).^13^
In our study, the median background noise measured was found 2.50 (IQR: 1.70-3.80) for Dixon sequence and that for STIR sequence was found 3.20 (IQR: 2.304.70) This quantitative parameter also highlights Dixon has better fat suppression technique in comparison to STIR technique. This statement was also supported by a similar study performed by Dalto VF et al.^14^
This study excluded all the patients under 18 years of age as well as those patients who required sedation during the examination. Thus, it is not possible to generalize this study results in patients below 18 years of age. Additionally, this study did not compare SI, CR and CNR in different clinical condition like disc degenerative disease, modic changes and marrow edema separately for evaluation of effectiveness of STIR and Dixon as fat suppression technique.
CONCLUSIONS
The quantative parameter suggests a better fat suppression by Dixon sequence which is comparable to other studies which shows higher degree of fat suppression along with less total scan time by Dixon sequence in comparison to that of STIR sequence for the evaluation of lumbar spine lesions.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1Lee S Choi DS Shin HS Baek HJ Choi HC Park SEFSE T 2-Weighted Two-Point Dixon Technique for Fat Suppression in the Lumbar Spine: Comparison With SPAIR Technique.Diagnostic and Interventional Radiology.2018 May 24317510.5152/dir.2018.1732029770772 PMC 5951208 · doi ↗ · pubmed ↗
- 2Kirchgesner T Perlepe V Michoux N Larbi A Berg BV Fat Suppression at 2D MR Imaging of the Hands: Dixon Method Versus CHESS Technique and STIR Sequence.European Journal of Radiology.2017 Apr 18940610.1016/j.ejrad.2017.01.01128267547 · doi ↗ · pubmed ↗
- 3Zadig P von Brandis E Ording Muller LS Tanturri De Horatio L Rosendahl K Avenarius DF Pediatric Whole-Body Magnetic Resonance Imaging: Comparison of STIR and T 2 Dixon Sequences in the Detection and Grading of High Signal Bone Marrow Changes.European Radiology.2023 Jul 33750455310.1007/s 00330-023-09413-636700955 PMC 10290001 · doi ↗ · pubmed ↗
- 4Brandao S Seixas D Ayres-Basto M Castro S Neto J Martins C Ferreira JC Parada F Comparing T 1-Weighted and T 2-Weighted Three-Point Dixon Technique With Conventional T 1-Weighted Fat-Saturation and Short-Tau Inversion Recovery (STIR) Techniques for the Study of the Lumbar Spine in a Short-Bore MRI Machine.Clinical radiology.2013 Nov 16811 e 6172310.1016/j.crad.2013.06.00423932678 · doi ↗ · pubmed ↗
- 5Maeder Y Dunet V Richard R Becce F Omoumi PB Bone Marrow Metastases: T 2-Weighted Dixon Spin-Echo Fat Images Can Replace T 1-Weighted Spin-Echo Images.Radiology.2018 Mar 28639485910.1148/radiol.201717032529095674 · doi ↗ · pubmed ↗
- 6Chen L Hu H Chen HH Chen W Wu Q Wu FY Xu XQ Usefulness of Two-Point Dixon T 2-Weighted Imaging in Thyroid-Associated Ophthalmopathy: Comparison With Conventional Fat Saturation Imaging in Fat Suppression Quality and Staging Performance.The British Journal of Radiology.2021 Feb 19411182020088410.1259/bjr.2020088433353397 PMC 7934300 · doi ↗ · pubmed ↗
- 7Bley TA Wieben O Francois CJ Brittain JH Reeder SB Fat and Water Magnetic Resonance Imaging.Journal of Magnetic Resonance Imaging.2010 Jan 31141810.1002/jmri.2189520027567 · doi ↗ · pubmed ↗
- 8Dagestad MH Toppe MK Kristoffersen PM Gjefsen E Andersen E Assmus J Vetti N Espeland A Dixon T 2 Imaging of Vertebral Bone Edema: Reliability and Comparison With Short-Tau Inversion Recovery.Acta Radiologica.2024 Mar 6532738310.1177/0284185122114613036560906 PMC 10964391 · doi ↗ · pubmed ↗