# Valve type and post-dilation impact on transprosthetic gradients in patients undergoing transcatheter aortic valve-in-valve procedure

**Authors:** Manuela Muratori, Laura Fusini, Gloria Tamborini, Paola Gripari, Sarah Ghulam Ali, Valentina Mantegazza, Anna Garlaschè, Francesco Doni, Andrea Baggiano, Francesco Cannata, Alberico Del Torto, Fabio Fazzari, Antonio Frappampina, Daniele Junod, Riccardo Maragna, Saima Mushtaq, Luigi Tassetti, Alessandra Volpe, Stefano Galli, Franco Fabbiocchi, Marco Gennari, Marco Agrifoglio, Antonio L Bartorelli, Federico De Marco, Mauro Pepi, Gianluca Pontone

PMC · DOI: 10.1093/ehjimp/qyaf048 · European Heart Journal. Imaging Methods and Practice · 2025-05-14

## TL;DR

This study examines how different valve types and balloon post-dilation affect heart valve replacement outcomes in patients with degenerated aortic valves.

## Contribution

The study identifies that self-expandable valves with balloon post-dilation result in lower transprosthetic gradients after valve-in-valve procedures.

## Key findings

- Self-expandable valves with balloon post-dilation had the lowest mean transprosthetic gradients at discharge.
- All groups showed improved hemodynamics over time, with no significant difference in 5-year mortality.
- Balloon post-dilation significantly reduced gradients compared to valves without post-dilation.

## Abstract

Valve-in-Valve transcatheter aortic valve replacement (ViV-TAVR) is an appealing treatment option for patients with degenerated aortic bioprosthetic valves. However, higher post-procedural transprosthetic gradients are more common after ViV-TAVR than after TAVR for native aortic valve stenosis. We sought to evaluate the impact of type of implanted valve and balloon post-dilation on echocardiographic results and mortality in ViV-TAVR patients.

One hundred and eleven consecutive patients were enrolled. A balloon-expandable valve, a self-expandable valve without balloon post-dilation, and a self-expandable valve with balloon post-dilation were performed in 35 (Group 1), 39 (Group 2), and 37 (Group 3) patients, respectively. All patients underwent comprehensive transthoracic echocardiography at baseline, discharge, and 6–12 months follow-up. Successful ViV-TAVR was performed in 110 patients (99%). Baseline transprosthetic gradients, left ventricular volumes, ejection fraction, and pulmonary artery systolic pressure were similar among groups. All groups experienced a significant reduction in post-procedural gradients at discharge and during the 6–12 months follow-up compared with baseline. At discharge, the lowest mean gradient was observed in Group 3 (12 ± 7 mmHg) compared with both Group 1 (20 ± 9 mmHg) and Group 2 (17 ± 8 mmHg, P = 0.001). This result was confirmed at 6–12 months follow-up (P = 0.012). Similar 5-year all-cause mortality was observed among groups (34%, 36%, 14%, respectively, P = 0.056).

In patients with failed surgical aortic prosthesis, ViV-TAVR is an effective treatment option associated with sustained improved haemodynamics regardless of transcatheter valve type and use of balloon post-dilation. However, self-expandable valves with balloon post-dilation showed lower transprosthetic gradients.

Graphical Abstract

## Linked entities

- **Diseases:** aortic valve stenosis (MONDO:0042981)

## Full-text entities

- **Diseases:** endocarditis (MESH:D004696), rupture (MESH:D012421), aortic valve degeneration (MESH:D000082862), valve malposition (MESH:D017760), Stenosis (MESH:D003251), calcification (MESH:D002114), thrombosis (MESH:D013927), aortic valve stenosis (MESH:D001024), death (MESH:D003643), leaks (MESH:D019559), valve failure (MESH:D006333), SVD (MESH:D006349), AR (MESH:D001022), atrial fibrillation (MESH:D001281), coronary obstruction (MESH:D000088442), paravalvular regurgitation (MESH:D008944)
- **Chemicals:** Edwards SAPIEN (-)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12076146/full.md

## Figures

2 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12076146/full.md

## References

31 references — full list in the complete paper: https://tomesphere.com/paper/PMC12076146/full.md

---
Source: https://tomesphere.com/paper/PMC12076146