# Aortoenteric Fistula: An Uncommon but Life-Threatening Complication of Pledget Use in Hiatal Hernia Repair

**Authors:** Jeet H Patel, Parag Brahmbhatt, Kalpit Devani, Jena Velji-Ibrahim

PMC · DOI: 10.7759/cureus.81432 · Cureus · 2025-03-29

## TL;DR

A rare case of a life-threatening aortoenteric fistula caused by a Teflon pledget used in hiatal hernia surgery is reported, highlighting the need for awareness and preventive measures.

## Contribution

This paper presents a rare complication of Teflon pledget erosion and offers insights into its prevention.

## Key findings

- A 74-year-old female developed an aortoenteric fistula due to erosion of a Teflon pledget.
- TEVAR and endoscopic removal of the pledget successfully treated the complication.
- The paper discusses causes of pledget erosion and possible countermeasures.

## Abstract

The standard treatment for hiatal hernia repair and gastroesophageal reflux disease (GERD) is laparoscopic Nissen fundoplication, in which a Teflon pledget (TP) is used to buttress the hiatal hernia. We present an extremely rare case in which a 74-year-old female developed a postoperative aortoenteric fistula due to erosion of a TP used during paraesophageal hernia repair. A computed tomography (CT) angiogram and esophagogastroduodenoscopy (EGD) confirmed the diagnosis. Subsequently, a thoracic endovascular aortic repair (TEVAR) was performed to treat the aortoenteric fistula, and the TP was removed using a raptor grasping device and scissors during a repeat EGD at a tertiary care center. In this article, we discuss different causes of TP erosion and possible countermeasures that may help prevent such complications.

## Linked entities

- **Diseases:** gastroesophageal reflux disease (MONDO:0007186)

## Full-text entities

- **Diseases:** Hiatal Hernia (MESH:D006551), GERD (MESH:D005764), Aortoenteric Fistula (MESH:D005402)

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC12037206/full.md

## Figures

2 figures with captions in the complete paper: https://tomesphere.com/paper/PMC12037206/full.md

## References

10 references — full list in the complete paper: https://tomesphere.com/paper/PMC12037206/full.md

---