MicroPort CRM considerations on Vega pacing lead performance
Andrea Vincon

Abstract
Genes, proteins, chemicals, diseases, species, mutations and cell lines named across the full text — each resolved to its canonical identifier and authoritative record.
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Taxonomy
Topics3D IC and TSV technologies · Electronic Packaging and Soldering Technologies · Electromagnetic Compatibility and Noise Suppression
Website: https://www.microport.com
We read with great interest the article by Özkartal et al.,^1^ which describes their single-centre experience with active fixation pacing leads. The authors report a high rate of complications associated with MicroPort’s Vega™ leads compared with other leads (Solia S™, Biotronik; Ingevity™, Boston-Scientific) and the rates published in MicroPort’s Product Performance Reports.
To accurately assess the performance of Vega™ leads, it is crucial to consider the experience gained from multicentre, international, controlled clinical investigations^2–5^ involving these pacing leads. From April 2018 to June 2023, 346 patients were implanted with 541 Vega™ leads (329 atrial and 212 ventricular) across 46 sites in 11 countries. The rate of Vega™ leads complications requiring intervention was 2.6% (1.9% from dislodgments and 0.7% from perforation\pericardial effusions) with a median follow-up of 22 months. These rates are consistent with existing literature.
Furthermore, the pooled multicentre data eliminate any bias that may arise from a single center^6,7^ and the practices of a limited number of healthcare providers.
Regarding the Product Performance Reports, MicroPort adheres to international standards, such as ISO 5841-2 (reporting of clinical performance of populations of pulse generators or leads). It is therefore justified and expected that the complication rates differ from those reported in the article, as they are based on different analysis methods. The authors noted that 45% of complications occurred within 30 days of implantation, while ISO 5841-2 states that acute lead complications must be reported separately in performance reports and not included in lead survival probability. This distinction is critical because acute complications may not be directly attributable to lead design but can result from various factors, including patient-specific anatomy, clinical conditions, implant techniques, and the experience of the implanters.
Additionally, more than 70% of the complications attributed to Vega™ in the article are linked to any pacing threshold increase > 2.0 V, even when capture and safety margins were maintained. However, ISO 5841-2 requires consideration of only significant increases where a 2:1 safety margin can no longer be achieved. Notably, the authors did not report most of the events for vigilance and, therefore, for inclusion in performance reports.
We acknowledge the work of Özkartal et al. and include it in our literature reviews. However, the data reported in the article should be evaluated in the context of a retrospective, monocentric cohort analysis. Their results are not in line with global data from larger populations that support the performance of Vega™ in line with the relevant literature and competitors’ leads.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1Özkartal T, Bergonti M, Caputo ML, Costantino J, Klersy C, Conte G. High lead-related complication rate with Micro Port Vega active fixation pacing leads. Europace 2024;26:euae 242.39288313 10.1093/europace/euae 242PMC 11448328 · doi ↗ · pubmed ↗
- 2Safety and performance of INVICTA ventricular DF 4 LEADS with active fixation (APOLLO) (https://www.clinicaltrials.gov/study/NCT 04590144)
- 3Clinical assessment of essential remote monitoring functions in pacemakers (CALLIOPE) (https://www.clinicaltrials.gov/study/NCT 05165095)
- 4Clinical investigation of ENO/TEO/OTO pacing system under MRI environment _APAC (CAPRI_APAC subgroup) (https://www.clinicaltrials.gov/study/NCT 03811691)
- 5Safety and electrical performances evaluation of Navigo Leads EQUIPPED with IS 4 connector (NAVIGATOR) (https://www.clinicaltrials.gov/study/NCT 03279484)
- 6Unverzagta S, Prondzinsky R, Peinemann F. Single-center trials tend to provide larger treatment effects than multicenter trials: a systematic review. J Clin Epidemiol 2013;66:1271–e 1280.23972520 10.1016/j.jclinepi.2013.05.016 · doi ↗ · pubmed ↗
- 7Dechartres A, Boutron I, Trinquart L, Charles P, Ravaud P. Single-center trials show larger treatment effects than multicenter trials: evidence from a meta-epidemiologic study. Ann Intern Med 2011;155:39–51.21727292 10.7326/0003-4819-155-1-201107050-00006 · doi ↗ · pubmed ↗