Considerations about the use of glucometers in glucose tolerance tests
Bernardo Alio Lavin-Gomez, Armando Raúl Guerra Ruíz

Abstract
Genes, proteins, chemicals, diseases, species, mutations and cell lines named across the full text — each resolved to its canonical identifier and authoritative record.
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Taxonomy
TopicsDiabetes Management and Research · Diabetes and associated disorders · Diet and metabolism studies
To the Editor,
We read with great interest the paper by Fabre-Estremera, B. et al. [1]. Congratulations for your article, which addresses a very relevant issue in the field of clinical biochemistry. The article provides a comparative study of POCT glucometers for professional use and measurement at the core laboratory to test for glucose tolerance in the diagnosis of prediabetes and diabetes. The paper also examines connectivity to the patient’s medical record and the associated cost reduction. Glucose test is a fundamental parameter in the diagnosis and management of patients with diabetes or glucose intolerance. Accurate glucose measurement is crucial in both in-hospital and out-of-hospital clinical activity. However, glucometers are often designed for self-monitoring rather than for professional use. Please, find below some critical considerations that may contribute to the debate about this issue.
- Pre-analytical sample handling: We find it noteworthy that, to prevent glycolysis during overload tests, containers with glycolysis inhibitors were not used at all time points. Current guidelines recommend using rapidly effective glycolysis inhibitor tubes. If this is not feasible, the tube containing the sample should be immediately placed in an ice-water slurry and centrifuged within 30 min from sample collection [2]. Failure to comply with these recommendations may lead to underestimate actual blood glucose concentrations due to post-collection glycolysis. This negatively affects comparative analysis of glucometers.
- Correlation between glucose measurements. The authors found statistically significant differences in fasting glucose concentrations, with glucometers showing higher values. Although these differences could be explained by the pre-analytical considerations mentioned in the section above and in the Discussion by the authors, the behavior of the samples on the regression analysis is remarkable. The regression plots and Passing-Bablok regressions in the Supplementary Material show a clear tendency of differences to increase significantly as glucose values increase in the core laboratory. This suggests a systematic bias at higher glucose values, with POCT yielding lower values as compared to the laboratory. Do you have any hypothesis that may explain this unexpected trend?
- Last, but not least, several studies suggest that abnormal hematocrit values may interfere with glucose readings on glucometers [3], [4], [5]. In general, low hematocrit values tend to lead to falsely elevated glucose readings, whereas high hematocrit values induce low glucose readings. In pediatric patients, it is recommended to establish a different range for the hematocrit based on age and sex. Thus, limits should range from 34 to 54 %, which may lead to significant differences [6]. However, most glucometers consider a constant hematocrit of 43 %, as recommended by the IFFC [7]. The same guidelines recommend using a constant factor of 11 % (F=1.11) for converting glucose in whole blood to glucose in plasma, a factor that is generally used by manufacturers.
Table 1 shows the results (unpublished data) obtained for glycemia in heparinized plasma measured in the core laboratory (Atellica^®^Solution-CH; Siemens-Healthineers, Erlangen, Germany) using the enzymatic method (glucose hexokinase), as compared to glycemia in heparinized whole blood measured on a professional glucometer that does not adjust the result to the real hematocrit of the patient and another commercially-available glucometer that concomitantly measures the hematocrit and performs separate conversion for each individual sample. When the glucose value was not adjusted for the actual hematocrit, a marked tendency was observed in the glucometer to yield higher glucose values at lower hematocrit values and vice versa (maximum rise of 23 % and maximum reductions of 19 %, respectively) despite the use of a constant conversion factor. The two glucometers complied with ISO 15197:2013 recommendations (at least 95 % of values within ±15 mg/dL for glycemias <100 mg/dL, and ±15 % for glycemias ≥100 mg/dL).
We agree with the authors that improvements in glucometers, both in analytical capabilities and in connectivity to electronic health records and full traceability, will increase their role in certain healthcare processes, reducing diagnostic and treatment delays, as well as analytical and clinical costs.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1Fabre-Estremera B Martínez-Chávez E Manzano Ocaña M Carcavilla UrquíA Morales Sánchez MDLÁ Pinilla Tejado I Uso de glucómetros durante la prueba de tolerancia oral a la glucosa en niños para el diagnóstico de prediabetes y diabetes. Estudio comparativo Adv Lab Med/Avances en Medicina de Laboratorio 2024519720410.1515/almed-2024-001738939199 PMC 11206185 · doi ↗ · pubmed ↗
- 2Sacks DB Arnold M Bakris GL Bruns DE Horvath AR LernmarkÅ Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus Clin Chem 2023698086810.1093/clinchem/hvad 08037473453 PMC 12376302 · doi ↗ · pubmed ↗
- 3Solnica B SkupieńJ Kuśnierz-Cabala B Słowińska-Solnica K Witek P Cempa A The effect of hematocrit on the results of measurements using glucose meters based on different techniques Clin Chem Lab Med 201250361510.1515/cclm.2011.77022047145 · doi ↗ · pubmed ↗
- 4Wada Y Nakamura T Kaneshige M Takahashi S Fujinaga H Tsukamoto K Evaluation of two glucose meters and interference corrections for screening neonatal hypoglycemia Pediatr Int 201557603710.1111/ped.1254325441549 · doi ↗ · pubmed ↗
- 5Demircik F Ramljak S Hermanns I Pfützner A Pfützner A Evaluation of hematocrit interference with My Star extra and seven competitive devices J Diabetes Sci Technol 20159262710.1177/193229681456579025549636 PMC 4604595 · doi ↗ · pubmed ↗
- 6FDA Blood glucose monitoring test systems for prescription point-of-care use: guidance for industry and food and drug administration staff U.S. Department of Health and Human Services 2020 Recuperado dehttps://www.fda.gov/regulatory-information/search-fda-guidance-documents/content-premarket-submissions-device-software-functions
- 7Burnett RW D’Orazio P Fogh-Andersen N Kuwa K Külpmann WR Larsson L Scientific division, working group on selective electrodes. IFCC recommendation on reporting results for blood glucose Clin Chim Acta 2001307205910.1016/s 0009-8981(01)00431-411369359 · doi ↗ · pubmed ↗
