# Ventricular fibrillation induced by atrial threshold search: a case report

**Authors:** David Sprenkeler, Ferry Hersbach, Ad Oomen, Clara van Ofwegen-Hanekamp, Mathias Meine

PMC · DOI: 10.1093/ehjcr/ytaf131 · European Heart Journal. Case Reports · 2025-03-15

## TL;DR

A man's implantable defibrillator triggered a dangerous heart rhythm during an automatic test, highlighting risks in device programming.

## Contribution

This case report highlights a rare but serious risk of ventricular fibrillation caused by automatic atrial threshold searches in ICDs.

## Key findings

- Ventricular fibrillation was triggered by an automatic atrial threshold search in a patient with a CRT-D device.
- Disabling Atrial Capture Management resolved the issue and prevented further arrhythmias.
- Newer ICD models reduce this risk by maintaining biventricular pacing during threshold tests.

## Abstract

Implantable cardioverter defibrillators (ICDs) have been proven to reduce the risk of sudden cardiac death from ventricular tachyarrhythmias. However, ICDs can sometimes induce malignant arrhythmias. We describe a case of ventricular fibrillation triggered by an automatic atrial threshold search.

A 72-year-old man presented after a syncopal episode. His medical history included moderate aortic regurgitation and a symptomatic second-degree atrioventricular (AV) block, for which he received a dual-chamber pacemaker in 2013, later upgraded to a CRT-D due to pacing-induced heart failure. ICD interrogation revealed an episode of ventricular fibrillation terminated by a shock. The arrhythmia started directly after an atrial threshold search. Extensive work-up did not reveal a cause of the arrhythmia, therefore, we considered it most likely that the atrial threshold test triggered the ventricular fibrillation. Atrial Capture Management was disabled, and the patient was discharged. No further ventricular arrhythmias or ICD therapies were observed.

Automatic threshold measurement algorithms are intended to ensure effective myocardial capture and enhance safety but can sometimes inadvertently cause arrhythmias. The underlying mechanism in this case may be related to the switch from biventricular to right ventricular (RV)-only pacing during Atrial Capture Management, which increases dispersion in repolarization, facilitating early afterdepolarizations and triggering polymorphic tachycardias. Notably, newer ICD models mitigate this risk by maintaining biventricular pacing during this test. This case underscores the need for careful programming and monitoring of ICD algorithms.

## Linked entities

- **Diseases:** atrioventricular block (MONDO:0000465), heart failure (MONDO:0005252)

## Full-text entities

- **Diseases:** aortic regurgitation (MESH:D001022), sudden cardiac death (MESH:D016757), Ventricular fibrillation (MESH:D014693), arrhythmia (MESH:D001145), heart failure (MESH:D006333), shock (MESH:D012769), syncopal episode (MESH:D013575), tachycardias (MESH:D013610), atrioventricular (AV) block (MESH:D054537)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC11941468/full.md

## Figures

5 figures with captions in the complete paper: https://tomesphere.com/paper/PMC11941468/full.md

## References

14 references — full list in the complete paper: https://tomesphere.com/paper/PMC11941468/full.md

---
Source: https://tomesphere.com/paper/PMC11941468