# Odontoid process type II and III fracture fixation using bone allograft screws versus cannulated screws: a biomechanical study

**Authors:** Emir Benca, Kenneth P. van Knegsel, Maximilian Pestel, Ivan Zderic, Jan Caspar, Lena Hirtler, Andreas Strassl, Dominic Gehweiler, Sonja Zehetmayer, Boyko Gueorguiev, Harald Widhalm, Reinhard Windhager, Peter Varga

PMC · DOI: 10.1007/s00402-025-05805-z · Archives of Orthopaedic and Trauma Surgery · 2025-03-22

## TL;DR

This study compares two methods for fixing odontoid fractures and finds that bone quality, not the method, mainly affects stability.

## Contribution

A biomechanical comparison of bone allograft screws and cannulated screws for odontoid fracture fixation.

## Key findings

- Both fixation methods restored similar levels of mechanical stability in odontoid fractures.
- Local bone quality, such as vBMD and cortical bone ratio, significantly influences biomechanical outcomes.
- Bone allograft screws showed slightly lower yield and ultimate load values compared to cannulated screws.

## Abstract

Fractures of the odontoid process are associated with high non-union rates, challenging treatment, and high incidence of screw-related complications. The aim of this study was to compare the biomechanical competence of a single biointegrative bone allograft screw versus two conventional cannulated screws for odontoid fracture fixation.

The odontoid process of intact C2 vertebral specimens was subjected to quasi-static loading until fracture. Specimens with an Anderson and d’Alonzo type II or III fracture (n = 47) were fixated with either two conventional cannulated screws or with a single bone allograft screw. The constructs were biomechanically tested to failure in the same fashion as in their intact state. Stiffness, yield, and ultimate load were evaluated. The results were adjusted by age, sex, volumetric bone mineral density (vBMD), and the cross-sectional area ratio of cortical bone to total bone measured at the junction of the odontoid process with the vertebral body (Ct.Ar/Tt.Ar).

Stiffness, yield and ultimate load were restored in the cannulated screws group by 44 ± 10%, 46 ± 7%, and 46 ± 5% and in the bone allograft group by 50 ± 12%, 30 ± 9%, and 34 ± 6% (mean ± SE). There were no significant differences between the groups regarding the three mechanical outcomes (0.104 ≤ p ≤ 0.223). Positive significant relation was found between vBMD and stiffness in each group (0.248 ≤ R²≤0.273, 0.018 ≤ p ≤ 0.038), as well as between Ct.Ar/Tt.Ar and stiffness (R²=0.218, p = 0.033), vBMD and ultimate load (R²=0.430, p = 0.001) and ultimate loadand vBMD (R²=0.315, p = 0.010) in the cannulated screws group.

The primary stability of odontoid fracture fixation is determined mainly by the quality of the local bone and independent of the fixation technique. From the biomechanical perspective, the lower mean values for the yield and ultimate load restored in the bone allograft group compared to the cannulated screws group should be compensated by lower peak load during the patient’s rehabilitation process.

## Full-text entities

- **Diseases:** Fractures of the odontoid process (MESH:D000092470), Anderson and d'Alonzo type II or III fracture (MESH:C535460), fracture (MESH:D050723)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC11929696/full.md