# Optimizing DSA parameters for enhanced radiation safety in interventional surgery

**Authors:** XunJin Zeng, Hao Wang, Guang Chen

PMC · DOI: 10.3389/fpubh.2025.1531240 · 2025-02-21

## TL;DR

This study identifies optimal settings for digital subtraction angiography to significantly reduce radiation exposure for patients and operators during interventional surgery.

## Contribution

The study introduces specific DSA parameter configurations that reduce radiation doses by up to 50% for patients and over 90% for operators using protective devices.

## Key findings

- Low-dose fluoroscopy mode reduced patient radiation dose by 50.8% and operator scattered dose by 25–34%.
- Raising the bed plate height from 80 to 110 cm reduced patient radiation dose by 45.5%.
- Proper use of protective devices reduced operator scatter radiation by over 10 times in some cases.

## Abstract

The study aims to establish a reliable method for reducing radiation dose by analyzing variations in radiation dose from digital subtraction machines (DSA).

The study investigates changes in bed plate height (80–110 cm), detection height (0–30 cm), visual field size (6 × 6 inches and 12 × 12 inches), and radiation doses affecting various body parts, including the lens, thyroid, chest, gonads, and lower limbs. Radiation doses were measured using Raysafe X2 dosimetry for patients and AT1123 meter for operators.

Compared to ordinary fluoroscopy, the low-dose fluoroscopy mode reduced the patient's radiation dose by 50.8% (from 13.2 to 6.5 mGy/min) and the operator's scattered radiation dose by 25–34% (lens dose reduced from 0.72 to 0.47 mGy). In photographic mode, the radiation dose was 3–4 times higher than in ordinary fluoroscopy (e.g., 53.9 vs. 13.2 mGy/min). Raising the bed plate height from 80 to 110 cm reduced the patient's direct radiation dose by 45.5% (from 24.2 to 13.2 mGy/min). The correct application of protective devices reduced the operator's scattering radiation by more than 10 times (e.g., gonads dose reduced from 4.07 to 0.41 mGy).

Selecting the appropriate bed plate (90–100 cm) and detector height (10–20 cm), along with an optimal visual field (6 × 6 inches), can effectively reduce radiation doses for both patients and operators. The proper use of protective devices in peripheral interventional surgery is crucial for reducing scatter radiation, with reductions exceeding 90% in some cases.

## Full-text entities

- **Species:** Homo sapiens (human, species) [taxon 9606]

## Figures

6 figures with captions in the complete paper: https://tomesphere.com/paper/PMC11885140/full.md

---
Source: https://tomesphere.com/paper/PMC11885140