# High-Power Short-Duration Posterior Wall Isolation in Addition to Pulmonary Vein Isolation in Persistent Atrial Fibrillation Ablation Using the New TactiFlex™ Ablation Catheter

**Authors:** Sergio Conti, Francesco Sabatino, Giulia Randazzo, Giuliano Ferrara, Antonio Cascino, Giuseppe Sgarito

PMC · DOI: 10.3390/jcdd11090294 · Journal of Cardiovascular Development and Disease · 2024-09-20

## TL;DR

This study shows that using the TactiFlex™ catheter for high-power short-duration ablation in treating persistent atrial fibrillation is safe and effective, with shorter procedure times.

## Contribution

The study evaluates the TactiFlex™ catheter for posterior wall isolation in addition to pulmonary vein isolation in persistent atrial fibrillation ablation.

## Key findings

- PVI and PWI were achieved in all patients using the TactiFlex™ catheter.
- Procedure and RF times were significantly shorter with TactiFlex™ compared to the TactiCath catheter.
- 76.8% of patients remained free from atrial arrhythmia at 6 months with no significant differences between groups.

## Abstract

Background: The TactiFlex™ ablation catheter, Sensor Enabled™ (Abbott, Minneapolis, MN, USA), is an open-irrigation radiofrequency (RF) ablation catheter with flexible tip technology. This catheter delivers high-power short-duration (HPSD) RF ablations and has been adopted for atrial fibrillation (AF) ablation. HPSD is well-established not only in pulmonary vein isolation (PVI) but also when targeting extra-pulmonary vein (PV) targets. This study aims to determine the safety, effectiveness, and acute outcomes of PVI plus posterior wall isolation (PWI) in patients with persistent atrial fibrillation (Pe-AF) using HPSD and the TactiFlex™ ablation catheter. Methods: Consecutive patients who underwent the ablation of Pe-AF in our centre between February 2023 and February 2024 were prospectively enrolled in the study. All patients underwent PVI plus PWI using TactiFlex™ and the HPSD strategy. The RF parameters were 50 W on all the PV segments and the roof, and within the posterior wall (PW). Left atrial mapping was performed with the EnSite X mapping system and the high-density multipolar Advisor HD Grid, Sensor Enabled™ mapping catheter. We compared the procedural data using HPSD with TactiFlex™ (n = 52) vs. a historical cohort of patients who underwent PVI plus PWI using HPSD settings and the TactiCath ablation catheter (n = 84). Results: Fifty-two consecutive patients were included in the study. PVI and PWI were achieved in all patients in the TactiFlex™ group. First-pass PVI was achieved in 97.9% of PVs (n = 195/199). PWI was obtained in all cases by delivering extensive RF lesions within the PW. There were no significant differences compared to the TactiCath group: first-pass PVI was achieved in 96.3% of PVs (n = 319/331). Adenosine administration revealed PV reconnection in 5.7% of patients, and two reconnections of the PW were documented. Procedure and RF time were significantly shorter in the TactiFlex™ group compared to the TactiCath group, 73.1 ± 12.6 vs. 98.5 ± 16.3 min, and 11.3 ± 1.5 vs. 23.5 ± 3.6 min, respectively, p < 0.001. The fluoroscopy time was comparable between both groups. No intraprocedural and periprocedural complications related to the ablation catheter were observed. Patients had an implantable loop recorder before discharge. At the 6-month follow-up, 76.8% of patients remained free from atrial arrhythmia, with no significant differences between groups. Conclusions: HPSD PVI plus PWI using the TactiFlex™ ablation catheter is effective and safe. Compared to a control group, the use of TactiFlex™ to perform HPSD PVI plus PWI is associated with a similar effectiveness but with a significantly shorter procedural and RF time.

## Linked entities

- **Diseases:** atrial fibrillation (MONDO:0004981)

## Full-text entities

- **Diseases:** AF (MESH:D001281), atrial arrhythmia (MESH:D001145)
- **Chemicals:** Adenosine (MESH:D000241), TactiFlex (-)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC11432285/full.md

## Figures

1 figure with captions in the complete paper: https://tomesphere.com/paper/PMC11432285/full.md

## References

40 references — full list in the complete paper: https://tomesphere.com/paper/PMC11432285/full.md

---
Source: https://tomesphere.com/paper/PMC11432285