# First-in-human cases and preclinical experience of a novel ICE catheter

**Authors:** Guangan Liu, Jingjing Wu, Fang Fang, Weipeng Zhao, Minmin Sun, Jihong Zhang, Simeng Liu, Mimi Wang, Feng Liu

PMC · DOI: 10.3389/fcvm.2024.1406470 · Frontiers in Cardiovascular Medicine · 2024-09-10

## TL;DR

A new ICE catheter was tested in humans for the first time, showing safety and effectiveness in cardiac imaging and procedures.

## Contribution

This is the first human use of a novel ICE catheter, demonstrating its practicality and image quality in clinical settings.

## Key findings

- The novel ICE catheter provided imaging quality comparable to existing commercial catheters.
- It successfully navigated complex cardiac structures without causing damage.
- Both patients were discharged without complications, indicating safety.

## Abstract

The primary objective of our study was to evaluate the first use of a novel intracardiac echocardiography (ICE) catheter in human subjects. This study aimed to assess its practicality, image clarity, and guidance role during electrophysiology procedures.

Two patients underwent procedures using the novel ICE catheter. Post-procedure evaluations were conducted by four operators, who assessed the imaging quality and overall performance of the catheter. Anatomical and blood test results were also analyzed to determine the safety and impact on internal cardiac structures.

Both patients were discharged one day after the procedure without any complications. The novel ICE catheter provided comparable imaging quality to existing commercial catheters. The catheter's advanced design allowed for detailed imaging at short distances, essential for accurate diagnosis and treatment planning. Moreover, it successfully navigated complex anatomical structures like the atrial septum and left atrial appendage.

These preliminary studies indicate that the novel ICE catheter achieves a level of safety and effectiveness comparable to previously available commercial catheters. The findings highlight its potential to meet current clinical needs, particularly in sophisticated anatomic interventions. Despite the prolonged thrombin time after anticoagulant administration, both types of ICE catheters were non-damaging to cardiac structures during routine operations. The study underscores the importance of using trans-septal large inner diameter sheaths to minimize complications when advancing the catheter into the left atrium.

## Full-text entities

- **Genes:** F2 (coagulation factor II, thrombin) [NCBI Gene 2147] {aka PT, RPRGL2, THPH1}
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC11419997/full.md

## Figures

7 figures with captions in the complete paper: https://tomesphere.com/paper/PMC11419997/full.md

## References

18 references — full list in the complete paper: https://tomesphere.com/paper/PMC11419997/full.md

---
Source: https://tomesphere.com/paper/PMC11419997