Entrapped surgical needle In the valveless AirSeal trocar: A case report and literature review
Ibrahim Can Aykanat, Mevlana Derya Balbay

TL;DR
A surgical needle was lost in a valveless AirSeal trocar during a robotic procedure and was found using retrograde inspection and X-ray.
Contribution
The paper introduces a method for locating lost instruments in valveless trocars using retrograde inspection and targeted radiography.
Findings
A missing needle was located within the AirSeal trocar using backward endoscopic inspection.
X-ray imaging of the trocar confirmed the presence of the needle.
Retrograde inspection can prevent unnecessary delays in surgery.
Abstract
Valveless trocars like AirSeal system are maintain a stable pneumoperitoneum and reduce instrument friction. A 65-year-old man's robotic radical cystectomy was complicated by a missing needle while using AirSeal system. The needle was detected via backward inspection using the endo camera inserted through the trocar, tip at its most distal end let the visualization of the needle within the air channels and confirmed with a trocar X-ray. Our findings suggest that retrograde inspection and targeted radiography of the trocar, prior to patient imaging, can be helpful in locating the lost needle and prevent prolongation of surgeries.
Genes, proteins, chemicals, diseases, species, mutations and cell lines named across the full text — each resolved to its canonical identifier and authoritative record.
Click any figure to enlarge with its caption.
Figure 1
Figure 2Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Taxonomy
TopicsAbdominal Surgery and Complications · Minimally Invasive Surgical Techniques · Anesthesia and Pain Management