# A Comparison of the Effectiveness of the McCoy Laryngoscope and the C-MAC D-Blade Video Laryngoscope in Obese Patients

**Authors:** Jung-Min Lee, Soo-Kyung Lee, Minsoo Jang, Minho Oh, Eun-Young Park

PMC · DOI: 10.3390/medicina60081285 · 2024-08-09

## TL;DR

This study compares two laryngoscopes for intubation in obese patients and finds the C-MAC D-blade may offer better hemodynamic stability.

## Contribution

The study provides a direct comparison of two laryngoscopes in obese patients with BMI ≥ 35.

## Key findings

- C-MAC group had shorter vocal cord exposure times and higher success rates within 5 seconds.
- C-MAC group showed lower intubation difficulty scale (IDS) values compared to the McCoy group.
- McCoy group required more lifting force during laryngoscopy, affecting hemodynamic variables.

## Abstract

Background and Objective: Obesity is associated with difficult or failed intubation attempts, making general anesthesia challenging for anesthesiologists to perform. The purpose of this study was to evaluate and compare the efficacy of a McCoy laryngoscope and a C-MAC D-blade video laryngoscope for intubation in obese patients with a body mass index (BMI) ≥ 35 kg/m2. Methods: In total, 104 patients were randomly assigned to be intubated with a McCoy (McCoy group) or C-MAC D-blade video laryngoscope (C-MAC group). The primary outcome was intubation time. The secondary outcomes were vocal cord exposure time, vocal cord passage time, proportion of successful intubation, mask ventilation scale, intubation difficulty scale (IDS), percentage of glottis opening (POGO) score, and hemodynamic variables. Results: Although the intubation time did not significantly differ, the C-MAC group showed shorter vocal cord exposure times and a higher rate of successful vocal cord exposure within 5 s. The IDS value was significantly lower in the C-MAC group than in the McCoy group. The proportion of patients who required an increase in lifting force during laryngoscopy was higher in the McCoy group than in the C-MAC group, which may explain the difference in MAP between the groups. Conclusions: Both the McCoy laryngoscope and the C-MAC D-blade video laryngoscope were useful during the intubation of obese patients. The C-MAC D-blade video laryngoscope might be more useful for obese patients in terms of hemodynamic stability.

## Linked entities

- **Diseases:** obesity (MONDO:0011122)

## Full-text entities

- **Diseases:** C-MAC (OMIM:211750), Obese (MESH:D009765)
- **Chemicals:** C-MAC (-)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Figures

3 figures with captions in the complete paper: https://tomesphere.com/paper/PMC11356276/full.md

---
Source: https://tomesphere.com/paper/PMC11356276