# Validation of a semi-quantitative scoring system and workflow for analysis of fluorescence quantification in companion animals

**Authors:** Ann S. Ram, Kathy Matuszewska, Charly McKenna, Jim Petrik, Michelle L. Oblak

PMC · DOI: 10.3389/fvets.2024.1392504 · Frontiers in Veterinary Science · 2024-07-31

## TL;DR

This paper validates a scoring system and workflow for measuring fluorescence in dogs during cancer surgery, enabling more objective assessments.

## Contribution

The study introduces a validated semi-quantitative scoring system and workflow for fluorescence quantification in veterinary surgical settings.

## Key findings

- Post hoc fluorescence quantification showed strong agreement with semi-quantitative scoring (k = 0.9734, p < 0.0001).
- Fluorescence intensity thresholds of 35–255 or 45–255 predicted fluorescence with high sensitivity and specificity (p < 0.05).
- Fluorescence intensity and quantification were strongly associated (p < 0.001).

## Abstract

Many commercially available near-infrared (NIR) fluorescence imaging systems lack algorithms for real-time quantifiable fluorescence data. Creation of a workflow for clinical assessment and post hoc analysis may provide clinical researchers with a method for intraoperative fluorescence quantification to improve objective outcome measures.

Scoring systems and verified image analysis are employed to determine the amount and intensity of fluorescence within surgical specimens both intra and postoperatively.

Lymph nodes from canine cancer patients were obtained during lymph node extirpation following peritumoral injection of indocyanine green (ICG). First, a semi-quantitative assessment of surface fluorescence was evaluated. Images obtained with a NIR exoscope were analysed to determine fluorescence thresholds and measure fluorescence amount and intensity.

Post hoc fluorescence quantification (threshold of Hue = 165–180, Intensity = 30–255) displayed strong agreement with semi-quantitative scoring (k = 0.9734, p < 0.0001). Fluorescence intensity with either threshold of 35–255 or 45–255 were significant predictors of fluorescence and had high sensitivity and specificity (p < 0.05). Fluorescence intensity and quantification had a strong association (p < 0.001).

The validation of the semi-quantitative scoring system by image analysis provides a method for objective in situ observation of tissue fluorescence. The utilization of thresholding for ICG fluorescence intensity allows post hoc quantification of fluorescence when not built into the imaging system.

## Linked entities

- **Chemicals:** indocyanine green (PubChem CID 5282412)
- **Diseases:** cancer (MONDO:0004992)

## Full-text entities

- **Diseases:** cancer (MESH:D009369)
- **Chemicals:** ICG (MESH:D007208)
- **Species:** Homo sapiens (human, species) [taxon 9606], Canis lupus familiaris (dog, subspecies) [taxon 9615]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC11322124/full.md

## Figures

2 figures with captions in the complete paper: https://tomesphere.com/paper/PMC11322124/full.md

## References

45 references — full list in the complete paper: https://tomesphere.com/paper/PMC11322124/full.md

---
Source: https://tomesphere.com/paper/PMC11322124