# Foam-based Left Atrial Appendage Closure (CLAAS) Device: Evaluation in a Chronic Canine Model

**Authors:** Aaron V. Kaplan, Carol Devellian, Jamie H. Kim, Jessica Rotschafer, Andy Levine, Robert J. Sommer, William A. Gray

PMC · DOI: 10.1016/j.jscai.2022.100555 · Journal of the Society for Cardiovascular Angiography & Interventions · 2022-12-01

## TL;DR

A new foam-based device for closing the left atrial appendage was tested in dogs and showed effective sealing without toxicity for up to 150 days.

## Contribution

The CLAAS device introduces a novel foam matrix for improved performance in left atrial appendage closure.

## Key findings

- CLAAS implants provided complete sealing of the left atrial appendage in all animals.
- No device-related toxicity or thrombus formation was observed in downstream organs.
- Implants maintained structural integrity and tissue integration over 150 days.

## Abstract

The Conformal Left Atrial Appendage System (CLAAS) is a transcatheter implant designed to occlude the left atrial appendage to prevent stroke in patients with nonvalvular atrial fibrillation. The implant incorporates a novel polyurethane polycarbonate-urea foam matrix to enhance device performance. The objective of this study was to characterize the thrombogenicity, sealing, implant integrity, local tissue response, and systemic toxicity of the implant in a healthy canine model.

Devices were evaluated in a chronic canine model using echocardiographic and fluoroscopic guidance (n = 8). Postprocedural echocardiographic evaluation was performed at case completion and at 2, 45, 90, and 150 days. Animals were sacrificed at 90 (n = 4) and 150 (n = 4) days. Necropsy included gross, x-ray, and histologic examination.

CLAAS implants were implanted in all animals successfully on the basis of fluoroscopic, intracardiac echo, and transthoracic echo criteria. Although 2 animals were noted to have leaks (1 at implantation and 1 at day 90 post procedure before sacrifice), both were of <5 mm. All left atrial appendages were completely sealed on gross examination. No device-related thrombus, pericardial effusion, or evidence of distal organ emboli was observed. All implants maintained their structural integrity and were well integrated within the tissue. There was no histologic evidence of toxicity noted in the downstream organs or tissues.

The CLAAS implant can be deployed providing a good seal for up to 150 days without signs of local or systemic toxicity in a canine model.

•Left atrial appendage closure is an emerging strategy for stroke reduction.•Conformal Left Atrial Closure System is a next-generation foam-based device.•Chronic canine studies showed complete closure at 90 and 150 days.

Left atrial appendage closure is an emerging strategy for stroke reduction.

Conformal Left Atrial Closure System is a next-generation foam-based device.

Chronic canine studies showed complete closure at 90 and 150 days.

## Linked entities

- **Diseases:** atrial fibrillation (MONDO:0004981), stroke (MONDO:0005098)
- **Species:** Canis lupus familiaris (taxon 9615)

## Full-text entities

- **Diseases:** atrial fibrillation (MESH:D001281), toxicity (MESH:D064420), stroke (MESH:D020521), leaks (MESH:D019559), Left Atrial Appendage (MESH:D059446), emboli (MESH:D020766), thrombus (MESH:D013927), pericardial effusion (MESH:D010490)
- **Species:** Canis lupus familiaris (dog, subspecies) [taxon 9615], Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC11308788/full.md

## Figures

4 figures with captions in the complete paper: https://tomesphere.com/paper/PMC11308788/full.md

## References

24 references — full list in the complete paper: https://tomesphere.com/paper/PMC11308788/full.md

---
Source: https://tomesphere.com/paper/PMC11308788