# Impact of acquisition volume on cone beam computed tomography imaging of marginal bone: an ex vivo study

**Authors:** Maurice Ruetters, Korallia Alexandrou, Holger Gehrig, Sinclair Awounvo, Ti-Sun Kim, Anna Felten, Christopher Lux, Sinan Sen

PMC · DOI: 10.2340/aos.v83.40494 · Acta Odontologica Scandinavica · 2024-04-25

## TL;DR

This study finds no clinically significant difference in measuring bone loss using high-dose or low-dose CBCT with different scan volumes.

## Contribution

The study provides empirical evidence on the equivalence of CBCT protocols for marginal bone loss assessment.

## Key findings

- High-dose and low-dose CBCT showed excellent agreement in measuring bone loss across different volumes.
- There was no substantial difference in accuracy between small and large acquisition volumes.
- Certainty scores varied significantly between high-dose and low-dose CBCT.

## Abstract

The current study explores whether there is a clinically relevant distinction in the measurement of marginal bone loss when comparing high-dose (HD) versus low-dose (LD) cone beam computed tomography (CBCT) protocols in small and large acquisition volumes.

CBCTs of four human cadaveric preparates were taken in HD and LD mode in two different fields of view 8 × 8 cm2 (LV) and 5 × 5 cm2 (SV). In total, 43 sites of 15 teeth were randomly chosen, and marginal bone loss was measured twice in all protocols at 43 sites of 15 teeth by one calibrated investigator. Bland-Altman plots and Lin’s concordance correlation coefficient (CCC) were calculated to assess the extent of agreement of the measurements. Additionally, the rater scored the certainty in each of the measurements.

For HD-CBCT CCC of measurements obtained using SV versus LV was 0.991. CCC of measurements obtained using SV versus LV of LD-CBCT was 0.963. Both CCC values indicated excellent agreement between the two volumes in both protocols.

CCC also indicated high intramodality correlation between HD-CBCT and LD-CBCT independent of the acquisition volume (0.963 – 0.992). Bland-Altman plots also indicated no substantial differences. Results of certainty scoring showed significant differences (p = 0.004 (LV), p < 0.001(SV)) between the LD and HD-CBCT.

Accuracy of measurements of bone loss shows no clinical noticeable effects depending on the CBCT volume in this ex vivo study. There appears to be no relevant advantage of SV over LV, neither in HD-CBCT nor in LD-CBCT and additionally no relevant advantage of HD versus LD in visualizing marginal bone loss.

## Linked entities

- **Species:** Homo sapiens (taxon 9606)

## Full-text entities

- **Diseases:** bone loss (MESH:D001847)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC11302637/full.md

## Figures

3 figures with captions in the complete paper: https://tomesphere.com/paper/PMC11302637/full.md

## References

17 references — full list in the complete paper: https://tomesphere.com/paper/PMC11302637/full.md

---
Source: https://tomesphere.com/paper/PMC11302637