# Chronic electrical stimulation with a peripheral suprachoroidal retinal implant: a preclinical safety study of neuroprotective stimulation

**Authors:** Carla J. Abbott, Penelope J. Allen, Chris E. Williams, Richard A. Williams, Stephanie B. Epp, Owen Burns, Ross Thomas, Mark Harrison, Patrick C. Thien, Alexia Saunders, Ceara McGowan, Caitlin Sloan, Chi D. Luu, David A. X. Nayagam

PMC · DOI: 10.3389/fcell.2024.1422764 · Frontiers in Cell and Developmental Biology · 2024-06-20

## TL;DR

A new implantable device for chronic retinal stimulation was tested in cats and found to be safe with minimal tissue response.

## Contribution

A minimally-invasive, implantable device for chronic retinal stimulation was developed and shown to be safe in preclinical trials.

## Key findings

- The device was well-tolerated with no adverse outcomes in all subjects.
- Electrode impedance remained stable over time, and retinal structure and function were preserved.
- Minimal inflammation was observed, with no retinal damage or scarring.

## Abstract

Extraocular electrical stimulation is known to provide neuroprotection for retinal cells in retinal and optic nerve diseases. Currently, the treatment approach requires patients to set up extraocular electrodes and stimulate potentially weekly due to the lack of an implantable stimulation device. Hence, a minimally-invasive implant was developed to provide chronic electrical stimulation to the retina, potentially improving patient compliance for long-term use. The aim of the present study was to determine the surgical and stimulation safety of this novel device designed for neuroprotective stimulation.

Eight normally sighted adult feline subjects were monocularly implanted in the suprachoroidal space in the peripheral retina for 9–39 weeks. Charge balanced, biphasic, current pulses (100 μA, 500 µs pulse width and 50 pulses/s) were delivered continuously to platinum electrodes for 3–34 weeks. Electrode impedances were measured hourly. Retinal structure and function were assessed at 1-, 2-, 4-, 6- and 8-month using electroretinography, optical coherence tomography and fundus photography. Retina and fibrotic thickness were measured from histological sections. Randomized, blinded histopathological assessments of stimulated and non-stimulated retina were performed.

All subjects tolerated the surgical and stimulation procedure with no evidence of discomfort or unexpected adverse outcomes. The device position was stable after a post-surgery settling period. Median electrode impedance remained within a consistent range (5–10 kΩ) over time. There was no change in retinal thickness or function relative to baseline and fellow eyes. Fibrotic capsule thickness was equivalent between stimulated and non-stimulated tissue and helps to hold the device in place. There was no scarring, insertion trauma, necrosis, retinal damage or fibroblastic response in any retinal samples from implanted eyes, whilst 19% had a minimal histiocytic response, 19% had minimal to mild acute inflammation and 28% had minimal to mild chronic inflammation.

Chronic suprathreshold electrical stimulation of the retina using a minimally invasive device evoked a mild tissue response and no adverse clinical findings. Peripheral suprachoroidal electrical stimulation with an implanted device could potentially be an alternative approach to transcorneal electrical stimulation for delivering neuroprotective stimulation.

## Full-text entities

- **Diseases:** retinal damage (MESH:D012164), inflammation (MESH:D007249), retinal and optic nerve diseases (MESH:D009901), trauma (MESH:D014947), necrosis (MESH:D009336)
- **Chemicals:** platinum (MESH:D010984)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC11222648/full.md

## Figures

13 figures with captions in the complete paper: https://tomesphere.com/paper/PMC11222648/full.md

## References

51 references — full list in the complete paper: https://tomesphere.com/paper/PMC11222648/full.md

---
Source: https://tomesphere.com/paper/PMC11222648