# Press-Fit Placement of a Rectangular Block Implant in the Resorbed Alveolar Ridge: Surgical and Biomechanical Considerations

**Authors:** Efthimios Gazelakis, Roy B. Judge, Joseph E. A. Palamara, Shiva Subramanian, Mohsin Nazir

PMC · DOI: 10.3390/bioengineering11060532 · Bioengineering · 2024-05-23

## TL;DR

A new implant placement technique for resorbed jawbones was tested in dogs and humans, showing good stability and success rates.

## Contribution

A novel surgical protocol for rectangular block implants in resorbed alveolar ridges is introduced with biomechanical insights.

## Key findings

- 16 out of 17 implant placements achieved full submergence and good stability without damaging the mandibular canal.
- Biomechanical success was linked to under-prepared osteotomy sites and stress distribution via implant design.
- One implant failed due to bone fracture, indicating the protocol's reliability with statistical confidence.

## Abstract

Rectangular Block Implant (RBIs) were manufactured, using computer-aided-design lathe turning, surface roughened with grit blasting and gamma irradiated. Implants were surgically placed into the resorbed edentulous mandibular ridges of both greyhound dogs (ex vivo and in vivo) and humans; the pooled total was 17 placements. The aim was to achieve mechanical stability and full implant submergence without damage to the mandibular canal and without bone fracture: fulfilment of all of these criteria was deemed to be a successful surgical outcome. Rectangular osteotomy sites were prepared with piezo surgical instrumentation. Sixteen implants were fully submerged and achieved good primary stability without bone fracture and without evidence of impingement of the mandibular canal. One implant placement was deemed a failure due to bone fracture: the event of a random successful outcome was rejected (p < 0.01 confidence, binomial analysis). Technique of placement yielded excellent mechanical retention: key biomechanical factors that emerged in this process included under preparation of the osteotomy site with the use of specifically designed trial-fit gauges, the viscoelastic property of the peri-implant bone, the flat faces and cornered edges of the block surfaces which enhance stress distribution and mechanical retention, respectively. It was concluded that the surgical protocol for the RBI placement in the resorbed alveolus is a predictable clinical procedure tailored to its specific, unique biomechanical profile.

## Linked entities

- **Species:** Homo sapiens (taxon 9606)

## Full-text entities

- **Diseases:** bone fracture (MESH:D050723)
- **Species:** Canis lupus familiaris (dog, subspecies) [taxon 9615], Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC11200368/full.md

## Figures

19 figures with captions in the complete paper: https://tomesphere.com/paper/PMC11200368/full.md

## References

39 references — full list in the complete paper: https://tomesphere.com/paper/PMC11200368/full.md

---
Source: https://tomesphere.com/paper/PMC11200368