# Performance of home-based self-tonometry (iCare HOME (TA022)) for measuring intraocular pressure among healthy and glaucoma patients

**Authors:** Anush Nayak, S Ve Ramesh, Neetha I R Kuzhuppilly, Vijaya H Pai, Aditya Chaitanya, Rashima Asokan, Aditya Chaitanya, Bhaskaran Mani, Sujatha V, Aditya Chaitanya

PMC · DOI: 10.12688/f1000research.123104.1 · 2023-02-02

## TL;DR

This study shows that the iCare HOME device allows patients to reliably measure their eye pressure at home, similar to a standard clinical method.

## Contribution

Demonstrates the reliability of home-based self-tonometry using iCare HOME in an Indian population.

## Key findings

- The iCare HOME device showed a mean difference of 0.83 mmHg compared to the Goldmann applanation tonometer.
- The intra-class correlation coefficient of the iCare HOME device was 0.997, indicating high reliability.
- Most participants successfully completed training and were certified to use the iCare HOME device.

## Abstract

Introduction: The purpose of this study was to compare iCare HOME (TA022) with Goldmann applanation tonometer and to evaluate the self-tonometry measurements among the Indian population.

Methods: Eligible patients underwent iCare HOME training through guided demonstration (verbal, pictorial, video) and practised self-tonometry measures using iCare HOME. Certification for independent iCare HOME measure was provided if first iCare HOME intraocular pressure (IOP) measurement fell within ± 5 mmHg of Goldmann applanation tonometer (GAT) measurement which was measured by the trained clinician (principal investigator). Certified participants underwent simulated home self-tonometry measurements using iCare HOME, and agreement with GAT IOP measurements was assessed.

Results: Seven of 83 participants (8.43%) failed to complete the study due to difficulty in performing the task, leading to non-certification. Patients who could use the iCare HOME had a mean age of 53 ± 15.55years (53% males; 46% females). Only one in 12 subjects did not qualify to use iCare HOME. The overall mean difference between iCare HOME and GAT was 0.83 mmHg (95%, 3.92 and -2.25). At various pressure ranges, 7-16 mmHg, 17-23 mmHg and >23 mmHg, the mean difference between iCare HOME and GAT was 1.22 mmHg (95%, 4.32 and -1.86), 0.77 mmHg (95%,3.69 and -2.19), -0.11 mmHg (95%, 2.52 and -2.74) respectively. The intra-class correlation coefficient of the iCare HOME device was 0.997(95% CI,0.995-0.998).

Conclusions: Patients were able to perform self- tonometry using iCare HOME with good reliability and safety. iCare HOME can be used to address the issue of difficulty in acquiring frequent and diurnal IOP measurements by patients doing self-tonometry from home.

## Linked entities

- **Diseases:** glaucoma (MONDO:0005041)

## Full-text entities

- **Diseases:** glaucoma (MESH:D005901)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Figures

5 figures with captions in the complete paper: https://tomesphere.com/paper/PMC11184277/full.md

---