# The comparison of white-to-white via triple person-times caliper measuring and machine-measuring in V4c implantable collamer lens implantation

**Authors:** Ting-Ting Dan, Tai-Xiang Liu, Zong-Ze Li, Ceng-Peng Liang, Fa-Yuan Li

PMC · DOI: 10.1038/s41598-024-64647-8 · 2024-06-16

## TL;DR

This study compares different methods for measuring eye dimensions before implanting a lens and finds that a manual caliper method is reliable and practical.

## Contribution

The study introduces triple person-times caliper as a practical and reliable method for ICL vault prediction.

## Key findings

- Triple person-times caliper measurements showed strong agreement with Pentacam HR.
- UBM measurements had the lowest correlation with other methods.
- The caliper method is more practical and less variable than other techniques.

## Abstract

This study aimed to compare the differences and characteristics of white-to-white (WTW) values obtained before V4c implantation using triple person-times caliper, IOL-Master 700, Pentacam HR, and UBM, and to assess their correlation with vaulting. A total of 930 myopia patients (1842 eyes) who were interested in undergoing ICL surgery were assessed before the procedure using various instruments. The WTW measurements were obtained using a triple person-times caliper, Pentacam HR, and IOL-Master 700, whereas the angle-to-angle (ATA) measurements were obtained using UBM. The size of the ICL was subsequently calculated using triple person-times caliper measurements. The vault of the ICL was assessed using Pentacam HR three months after the surgery. The WTW was determined to be 11.30 ± 0.29 mm, 11.43 ± 0.29 mm, and11.86 ± 0.38 mm, respectively, using the triple person-times caliper, Pentacam HR, and IOL-Master 700. The measurement of ATA was 11.57 ± 0.51 mm, as done by UBM. The ICL vault was measured to be 400.97 ± 198.46 µm when examined with Pentacam HR three monthsafter the procedure. The linear regression analyses of ICL size and WTW of triple person-times caliper, ICL vault and WTW were (R = 0.703, p < 0.001; R = 0.0969, p < 0.001) respectively. The highest correlation was found between IOL-Master and Pentacam HR (r = 0.766, p = 0.000). The lowest correlation was found between UBM and Pentacam HR (r = 0.358, p = 0.002). Bland–Altman analysis showed that the 95% limits of agreement (LoA) were the triple person-times caliper and Pentacam HR (– 0.573, 0.298) and the triple person-times caliper and UBM (– 1.15, – 0.605). This indicated a strong agreement between the triple person-times caliper and Pentacam HR and a lack of agreement between the triple person-times caliper and UBM. Triple person-times caliper measurements offer excellent maneuverability, practicality, and reliable outcomes for determining ICL vaults. Measurements obtained using the triple-person caliper were less differece than those obtained using the Pentacam HR.

## Full-text entities

- **Diseases:** myopia (MESH:D009216)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Figures

4 figures with captions in the complete paper: https://tomesphere.com/paper/PMC11180651/full.md

---
Source: https://tomesphere.com/paper/PMC11180651